Midmark® 1000 Checking Foot Control Test and Repair

Midmark® 1000 Foot Control

To check a leaking foot controller

1.Check for hole or puncture in tubing.

2.Check springs, if compressed, stretch them out and reinstall.

3.Use soapy water to check o-rings for leaks.

4.Check tri-block diaphragm to see if it has a puncture. Tri-block is located behind the delivery panel. Refer to: Access Procedures and Checking for Leaks.