Power Supply 153941 (002-2018-XX) Replacing Fuses Test and Repair
To prevent the possibility of electrical shock, always disconnect electrical mains supply to the Power Supply before removing the cover. Failure to comply with these instructions could result in severe personal injury or death.
To Replace Fuses
Lift straight up to remove button type fuses.
1.Disconnect power.
2.Remove cover and top. Reference Accessing the Power Suppl
3.Extract fuse(s) to be replaced.
4.Install new fuse(s).
5.Install cover and top then secure with screws and lock washers.
6.Reconnect powercord to the Power Supply.