Midmark® Dental Halogen Light Beam Focus Adjustments

Mounted To

Ceiling, Chair, Track, Universal and Wall/Cabinet
Serial Number NW, RE and V Serial Number Prefixes

Midmark® Dental Lights Light Beam Focus Adjustments

WARNING

Refer to: Midmark Dental Light User Guide for complete instructions on operating the light. Failure to do so could result in personal injury.

Perform an operational test on the dental light after repair is completed to confirm repair was properly made and that all malfunctions were repaired.

To Adjust Light Beam Focus

This adjustment affects the sharpness of the light pattern at a specified distance. The light’s focus is factory preset for a distance of 27 in. (68.6 cm) from the front surface of the plastic shield to patient (target). However, the focus can be adjusted per user’s preference for any distance between 21 to 39 in. (53.3 to 99.0 cm).

1.Position plastic shield (1) of lighthead at any distance between 21 to 39 in. (53.3 to 99.0 cm) from a target (2) (wall, chair, or back section) and turn on light.

2.Using a screwdriver, pry focus cap (3) off back of lighthead.

3.Using a 1/8 in. Allen Wrench (provided), adjust focus screw (4) as necessary to achieve desired pattern. The suggested pattern is a 3 x 8 in. (8 x 20 cm) pattern, oval in shape, with feathered edges.

4.Install focus cap (3) on back of lighthead.