M3 Sterilizer General System Error Codes: C630 to C675 Troubleshooting
Items Not Sterile message will be displayed if an error occurs in any of the following modes: Close mode, Heat Up Mode, or Sterilize Mode. Otherwise, message will not appear. This message is intended to indicate to the user that the items in the chamber may or may not be sterile, and must be reprocessed.
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C630: POWER UP MODE PRESSURE LOW ITEMS NOT STERILE TURN POWER OFF/ON During Heatup Mode, the first vent valve open operation has to occur within 10 minutes. |
Low water in reservoir | Check water level in reservoir. | Fill reservoir with distilled water. |
Low water in reservoir not caught by a C23x error. | Water Level / External Tank Display Screen. | Replace water level sensor. | |
Water pump not primed. | Water pump priming | Prime water pump. | |
Steam leakage. | Repair steam leak. | ||
Chamber temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Clean or replace chamber temperature sensor. | |
Boiler temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Replace boiler temperature sensor. | |
Boiler not functioning properly. | Boiler Test | Replace boiler, if necessary. | |
Pressure transducer tubing damaged or disconnected. | Inspect copper & neoprene tubing from chamber to pressure transducer. | Repair or replace tubing, if necessary. | |
PC Board malfunctioning. | Verify that there are no other ways that pressure could not be transmitted from chamber to pressure transducer. | Replace PC board, if necessary. | |
Defective check valve in water pump. | Check for large air pocket forming in tubing, that moves from the boiler to water pump. | Replace water pump. | |
Pressure transducer port in chamber blocked | Check for obstruction. | Remove obstruction. | |
Photo interrupter obstructed. | Photo Interrupter | Clear obstruction and clean. Verify hardware is working back to processor. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C631: SELECT CYCLE MODE PRESSURE LOW ITEMS NOT STERILE TURN POWER OFF/ON During Heatup Mode, the first vent valve open operation has to occur within 10 minutes. |
Low water in reservoir | Check water level in reservoir. | Fill reservoir with distilled water. |
Low water in reservoir not caught by a C23x error. | Water Level / External Tank Display Screen. | Replace water level sensor. | |
Water pump not primed. | Water pump priming | Prime water pump. | |
Steam leakage. | Repair steam leak. | ||
Chamber temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Clean or replace chamber temperature sensor. | |
Boiler temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Replace boiler temperature sensor. | |
Boiler not functioning properly. | Boiler Test | Replace boiler, if necessary. | |
Pressure transducer tubing damaged or disconnected. | Inspect copper & neoprene tubing from chamber to pressure transducer. | Repair or replace tubing, if necessary. | |
PC Board malfunctioning. | Verify that there are no other ways that pressure could not be transmitted from chamber to pressure transducer. | Replace PC board, if necessary. | |
Defective check valve in water pump. | Check for large air pocket forming in tubing, that moves from the boiler to water pump. | Replace water pump. | |
Pressure transducer port in chamber blocked | Check for obstruction. | Remove obstruction. | |
Photo interrupter obstructed. | Photo Interrupter | Clear obstruction and clean. Verify hardware is working back to processor. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C633: HEAT UP MODE PRESSURE LOW ITEMS NOT STERILE TURN POWER OFF/ON During Heatup Mode, the first vent valve open operation has to occur within 10 minutes. |
Low water in reservoir | Check water level in reservoir. | Fill reservoir with distilled water. |
Low water in reservoir not caught by a C23x error. | Water Level / External Tank Display Screen | Replace water level sensor. | |
Water pump not primed. | Water pump priming | Prime water pump. | |
Steam leakage. | Repair steam leak. | ||
Chamber temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Clean or replace chamber temperature sensor. | |
Boiler temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Replace boiler temperature sensor. | |
Boiler not functioning properly. | Boiler Test | Replace boiler, if necessary. | |
Pressure transducer tubing damaged or disconnected. | Inspect copper & neoprene tubing from chamber to pressure transducer. | Repair or replace tubing, if necessary. | |
PC Board malfunctioning. | Verify that there are no other ways that pressure could not be transmitted from chamber to pressure transducer. | Replace PC board, if necessary. | |
Defective check valve in water pump. | Check for large air pocket forming in tubing, that moves from the boiler to water pump. | Replace water pump. | |
Pressure transducer port in chamber blocked | Check for obstruction. | Remove obstruction. | |
Photo interrupter obstructed. | Photo Interrupter | Clear obstruction and clean. Verify hard-ware is working back to processor. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C636: OPEN MODE PRESSURE LOW ITEMS NOT STERILE TURN POWER OFF/ON During Heatup Mode, the first vent valve open operation has to occur within 10 minutes. |
Low water in reservoir | Check water level in reservoir. | Fill reservoir with distilled water. |
Low water in reservoir not caught by a C23x error. | Water Level / External Tank Display Screen | Replace water level sensor. | |
Water pump not primed. | Water pump priming | Prime water pump. | |
Steam leakage. | Repair steam leak. | ||
Chamber temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Clean or replace chamber temperature sensor. | |
Boiler temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Replace boiler temperature sensor. | |
Boiler not functioning properly. | Boiler Test | Replace boiler, if necessary. | |
Pressure transducer tubing damaged or disconnected. | Inspect copper & neoprene tubing from chamber to pressure transducer. | Repair or replace tubing, if necessary. | |
PC Board malfunctioning. | Verify that there are no other ways that pressure could not be transmitted from chamber to pressure transducer. | Replace PC board, if necessary. | |
Defective check valve in water pump. | Check for large air pocket forming in tubing, that moves from the boiler to water pump. | Replace water pump. | |
Pressure transducer port in chamber blocked | Check for obstruction. | Remove obstruction. | |
Photo interrupter obstructed. | Photo Interrupter |
Clear obstruction and clean. Verify hardware is working back to processor.
|
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C637: DRY MODE PRESSURE LOW ITEMS NOT STERILE TURN POWER OFF/ON During Heatup Mode, the first vent valve open operation has to occur within 10 minutes. |
Low water in reservoir | Check water level in reservoir. | Fill reservoir with distilled water. |
Low water in reservoir not caught by a C23x error. | Water Level / External Tank Display Screen | Replace water level sensor. | |
Water pump not primed. | Water pump priming. | Prime water pump. | |
Steam leakage. | Repair steam leak. | ||
Chamber temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Clean or replace chamber temperature sensor. | |
Boiler temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Replace boiler temperature sensor. | |
Boiler not functioning properly. | Boiler Test | Replace boiler, if necessary. | |
Pressure transducer tubing damaged or disconnected. | Inspect copper & neoprene tubing from chamber to pressure transducer. | Repair or replace tubing, if necessary. | |
PC Board malfuctioning. | Varify that there are no other ways that pressure could not be transmitted from chamber to pressure transducer. | Replace PC board, if necessary. | |
Defective check valve in water pump. | Check for large air pocket forming in tubing, that moves from the boiler to water pump. | Replace water pump. | |
Pressure transducer port in chamber blocked | Check for obstruction. | Remove obstruction. | |
Photo interrupter obstructed. | Photo Interrupter | Clear obstruction and clean. Verify hardware is working back to processor. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C638: COMPLETE MODE PRESSURE LOW ITEMS NOT STERILE TURN POWER OFF/ON During Heatup Mode, the first vent valve open operation has to occur within 10 minutes. |
Low water in reservoir | Check water level in reservoir. | Fill reservoir with distilled water. |
Low water in reservoir not caught by a C23x error. | Water Level / External Tank Display Screen | Replace water level sensor. | |
Water pump not primed. | Water pump priming. | Prime water pump. | |
Steam leakage. | Repair steam leak. | ||
Chamber temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Clean or replace chamber temperature sensor. | |
Boiler temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Replace boiler temperature sensor. | |
Boiler not functioning properly. | Boiler Test | Replace boiler, if necessary. | |
Pressure transducer tubing damaged or disconnected. | Inspect copper & neoprene tubing from chamber to pressure transducer. | Repair or replace tubing, if necessary. | |
PC Board malfuctioning. | Varify that there are no other ways that pressure could not be transmitted from chamber to pressure transducer. | Replace PC board, if necessary. | |
Defective check valve in water pump. | Check for large air pocket forming in tubing, that moves from the boiler to water pump. | Replace water pump. | |
Pressure transducer port in chamber blocked | Check for obstruction. | Remove obstruction. | |
Photo interrupter obstructed. | Photo Interrupter | Clear obstruction and clean. Verify hardware is working back to processor. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C641: SELECT MODE PRESSURE HIGH ITEMS NOT STERILE PUSH STOP TO RESTART The gauge pressure in the chamber is greater than 2.0 PSIg (13.8 kPag) when it should be near zero. |
Steam not being released from chamber. | Check for obstruction in plastic tube connecting unit to external condensing tank. | Unkink plastic tube, remove obstruction, or replace. |
Check copper coil in external condensing tank for blockage or damage. | Clear obstruction or replace copper coil. | ||
Check vent port in chamber for blockage. Check filter for blockage. | Clear obstruction in vent port. Clean or replace filter. | ||
Check vent valve for blockage. Put unit into Service Diagnostics and check vent valve operation. | Clean or replace vent valve. | ||
Pressure is being read abnormally high by pressure transducer. | Chamber Pressure Display Screen | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C642: CLOSE MODE PRESSURE HIGH ITEMS NOT STERILE PUSH STOP TO RESTART The gauge pressure in the chamber is greater than 2.0 PSIg (13.8 kPag) when it should be near zero. |
Steam not being released from chamber. | Check for obstruction in plastic tube connecting unit to external condensing tank. | Unkink plastic tube, remove obstruction, or replace. |
Check copper coil in external condensing tank for blockage or damage. | Clear obstruction or replace copper coil. | ||
Check vent port in chamber for blockage. Check filter for blockage. | Clear obstruction in vent port. Clean or replace filter. | ||
Check vent valve for blockage. Put unit into Service Diagnostics and check vent valve operation. | Clean or replace vent valve. | ||
Pressure is being read abnormally high by pressure transducer. | Chamber Pressure Display Screen | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C645: VENT MODE PRESSURE HIGH ITEMS NOT STERILE PUSH STOP TO RESTART The gauge pressure in the chamber is greater than 2.0 PSIg (13.8 kPag) when it should be near zero. |
Steam not being released from chamber. | Check for obstruction in plastic tube connecting unit to external condensing tank. | Unkink plastic tube, remove obstruction, or replace. |
Check copper coil in external condensing tank for blockage or damage. | Clear obstruction or replace copper coil. | ||
Check vent port in chamber for blockage. Check filter for blockage. | Clear obstruction in vent port. Clean or replace filter. | ||
Check vent valve for blockage. Put unit into Service Diagnostics and check vent valve operation. | Clean or replace vent valve. | ||
Pressure is being read abnormally high by pressure transducer. | Chamber Pressure Display Screen | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C646: OPEN MODE PRESSURE HIGH ITEMS NOT STERILE PUSH STOP TO RESTART The gauge pressure in the chamber is greater than 2.0 PSIg (13.8 kPag) when it should be near zero. |
Steam not being released from chamber. | Check for obstruction in plastic tube connecting unit to external condensing tank. | Unkink plastic tube, remove obstruction, or replace. |
Check copper coil in external condensing tank for blockage or damage. | Clear obstruction or replace copper coil. | ||
Check vent port in chamber for blockage. Check filter for blockage. | Clear obstruction in vent port. Clean or replace filter. | ||
Check vent valve for blockage. Put unit into Service Diagnostics and check vent valve operation. | Clean or replace vent valve. | ||
Pressure is being read abnormally high by pressure transducer. | Chamber Pressure Display Screen | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C647: DRY MODE PRESSURE HIGH ITEMS NOT STERILE PUSH STOP TO RESTART The gauge pressure in the chamber is greater than 2.0 PSIg (13.8 kPag) when it should be near zero. |
Steam not being released from chamber. | Check for obstruction in plastic tube connecting unit to external condensing tank. | Unkink plastic tube, remove obstruction, or replace. |
Check copper coil in external condensing tank for blockage or damage. | Clear obstruction or replace copper coil. | ||
Check vent port in chamber for blockage. Check filter for blockage. | Clear obstruction in vent port. Clean or replace filter. | ||
Check vent valve for blockage. Put unit into Service Diagnostics and check vent valve operation. | Clean or replace vent valve. | ||
Pressure is being read abnormally high by pressure transducer. | Chamber Pressure Display Screen | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C648: COMPLETE MODE PRESSURE HIGH ITEMS NOT STERILE PUSH STOP TO RESTART The gauge pressure in the chamber is greater than 2.0 PSIg (13.8 kPag) when it should be near zero. |
Steam not being released from chamber. | Check for obstruction in plastic tube connecting unit to external condensing tank. | Unkink plastic tube, remove obstruction, or replace. |
Check copper coil in external condensing tank for blockage or damage. | Clear obstruction or replace copper coil. | ||
Check vent port in chamber for blockage. Check filter for blockage. | Clear obstruction in vent port. Clean or replace filter. | ||
Check vent valve for blockage. Put unit into Service Diagnostics and check vent valve operation. | Clean or replace vent valve. | ||
Pressure is being read abnormally high by pressure transducer. | Chamber Pressure Display Screen. | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C660: POWER UP MODE PRESSURE HARDWARE ITEMS NOT STERILE TURN POWER OFF/ON A/D converter reports a gauge pressure input value outside the normal limits. |
Electrical interruption. | If error is intermittent. | None. Ignore code. |
PC Board malfunctioning. | If error is persistent, then a problem exist. | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C661: SELECT MODE PRESSURE HARDWARE ITEMS NOT STERILE TURN POWER OFF/ON A/D converter reports a gauge pressure input value outside the normal limits. |
Electrical interruption. | If error is intermittent. | None. Ignore code. |
PC Board malfunctioning. | If error is persistent, then a problem exist. | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C662: CLOSE MODE PRESSURE HARDWARE ITEMS NOT STERILE TURN POWER OFF/ON A/D converter reports a gauge pressure input value outside the normal limits. |
Electrical interruption. | If error is intermittent. | None. Ignore code. |
PC Board malfunctioning. | If error is persistent, then a problem exist. | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C663: HEAT UP MODE PRESSURE HARDWARE ITEMS NOT STERILE TURN POWER OFF/ON A/D converter reports a gauge pressure input value outside the normal limits. |
Electrical interruption. | If error is intermittent. | None. Ignore code. |
PC Board malfunctioning. | If error is persistent, then a problem exist. | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C664: STERILIZE MODE PRESSURE HARDWARE ITEMS NOT STERILE TURN POWER OFF/ON A/D converter reports a gauge pressure input value outside the normal limits. |
Electrical interruption. | If error is intermittent. | None. Ignore code. |
PC Board malfunctioning. | If error is persistent, then a problem exist. | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C665: VENT MODE PRESSURE HARDWARE ITEMS NOT STERILE TURN POWER OFF/ON A/D converter reports a gauge pressure input value outside the normal limits. |
Electrical interruption. | If error is intermittent. | None. Ignore code. |
PC Board malfunctioning. | If error is persistent, then a problem exist. | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C666: OPEN MODE PRESSURE HARDWARE ITEMS NOT STERILE TURN POWER OFF/ON A/D converter reports a gauge pressure input value outside the normal limits. |
Electrical interruption. | If error is intermittent. | None. Ignore code. |
PC Board malfunctioning. | If error is persistent, then a problem exist. | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C667: DRY MODE PRESSURE HARDWARE ITEMS NOT STERILE TURN POWER OFF/ON A/D converter reports a gauge pressure input value outside the normal limits. |
Electrical interruption. | If error is intermittent. | None. Ignore code. |
PC Board malfunctioning. | If error is persistent, then a problem exist. | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C668: COMPLETE MODE PRESSURE HARDWARE ITEMS NOT STERILE TURN POWER OFF/ON A/D converter reports a gauge pressure input value outside the normal limits. |
Electrical interruption. | If error is intermittent. | None. Ignore code. |
PC Board malfunctioning. | If error is persistent, then a problem exist. | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C670: POWER UP MODE PRESSURE OVER LIMIT ITEMS NOT STERILE TURN POWER OFF/ON Gauge chamber pressure greater than 40 PSI (275.8 kPa) |
Steam not being released from chamber. | Check for obstruction in plastic tube connecting unit to external condensing tank | Unkink plastic tube, remove obstruction, or replace. |
Check copper coil in external condensing tank for blockage or damage. | Clear obstruction or replace copper coil. | ||
Check vent port in chamber for blockage. Check filter for blockage. | Clear obstruction in vent port. Clean or replace filter. | ||
Check vent valve for blockage. Put unit into Service Diagnostics and check vent valve operation. | Clean or replace vent valve. | ||
Chamber temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Clean or replace chamber temperature sensor. | |
Boiler temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Replace boiler temperature sensor. | |
PC Board is not processing temperature input(s) properly. | Verify temperature sensors are working properly. | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C671: SELECT MODE PRESSURE OVER LIMIT ITEMS NOT STERILE TURN POWER OFF/ON Gauge chamber pressure greater than 40 PSI (275.8 kPa) |
Steam not being released from chamber. | Check for obstruction in plastic tube connecting unit to external condensing tank | Unkink plastic tube, remove obstruction, or replace. |
Check copper coil in external condensing tank for blockage or damage. | Clear obstruction or replace copper coil. | ||
Check vent port in chamber for blockage. Check filter for blockage. | Clear obstruction in vent port. Clean or replace filter. | ||
Check vent valve for blockage. Put unit into Service Diagnostics and check vent valve operation. | Clean or replace vent valve. | ||
Chamber temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Clean or replace chamber temperature sensor. | |
Boiler temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Replace boiler temperature sensor. | |
PC Board is not processing temperature input(s) properly. | Verify temperature sensors are working properly. | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C673: HEAT UP MODE PRESSURE OVER LIMIT ITEMS NOT STERILE TURN POWER OFF/ON Gauge chamber pressure greater than 40 PSI (275.8 kPa) |
Steam not being released from chamber. | Check for obstruction in plastic tube connecting unit to external condensing tank | Unkink plastic tube, remove obstruction, or replace. |
Check copper coil in external condensing tank for blockage or damage. | Clear obstruction or replace copper coil. | ||
Check vent port in chamber for blockage. Check filter for blockage. | Clear obstruction in vent port. Clean or replace filter. | ||
Check vent valve for blockage. Put unit into Service Diagnostics and check vent valve operation. | Clean or replace vent valve. | ||
Chamber temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Clean or replace chamber temperature sensor. | |
Boiler temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Replace boiler temperature sensor. | |
PC Board is not processing temperature input(s) properly. | Verify temperature sensors are working properly. | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C674: STERILIZE MODE PRESSURE OVER LIMIT ITEMS NOT STERILE TURN POWER OFF/ON Gauge chamber pressure greater than 40 PSI (275.8 kPa) |
Steam not being released from chamber. | Check for obstruction in plastic tube connecting unit to external condensing tank | Unkink plastic tube, remove obstruction, or replace. |
Check copper coil in external condensing tank for blockage or damage. | Clear obstruction or replace copper coil. | ||
Check vent port in chamber for blockage. Check filter for blockage. | Clear obstruction in vent port. Clean or replace filter. | ||
Check vent valve for blockage. Put unit into Service Diagnostics and check vent valve operation. | Clean or replace vent valve. | ||
Chamber temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Clean or replace chamber temperature sensor. | |
Boiler temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Replace boiler temperature sensor. | |
PC Board is not processing temperature input(s) properly. | Verify temperature sensors are working properly. | Replace PC board, if necessary. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C675: VENT MODE PRESSURE OVER LIMIT ITEMS NOT STERILE TURN POWER OFF/ON Gauge chamber pressure greater than 40 PSI (275.8 kPa) |
Steam not being released from chamber. | Check for obstruction in plastic tube connecting unit to external condensing tank | Unkink plastic tube, remove obstruction, or replace. |
Check copper coil in external condensing tank for blockage or damage. | Clear obstruction or replace copper coil. | ||
Check vent port in chamber for blockage. Check filter for blockage. | Clear obstruction in vent port. Clean or replace filter. | ||
Check vent valve for blockage. Put unit into Service Diagnostics and check vent valve operation. | Clean or replace vent valve. | ||
Chamber temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Clean or replace chamber temperature sensor. | |
Boiler temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Replace boiler temperature sensor. | |
PC Board is not processing temperature input(s) properly. | Verify temperature sensors are working properly. | Replace PC board, if necessary. |