M3 Sterilizer General System Error Codes: C981 to C987 Troubleshooting
Items Not Sterile message will be displayed if an error occurs in any of the following modes: Close mode, Heat Up Mode, or Sterilize Mode. Otherwise, message will not appear. This message is intended to indicate to the user that the items in the chamber may or may not be sterile, and must be reprocessed.
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C981: SELECT MODE BOILER HI-LIMIT OPEN ITEMS NOT STERILE TURN POWER OFF/ON Boiler overheat thermostat has opened for at least 0.25 seconds & needs to be manually reset. |
Water pump not primed. | Water pump priming instructions. | Prime water pump. |
Boiler temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Replace boiler temperature sensor. | |
Boiler overheat thermostat disconnected at PC board. | Check wire connections at J3 on PC board. | Secure wire connections. | |
Loose or broken wire connections at boiler overheat thermostat. | Check wire connections at boiler overheat thermostat. | Secure wire connections. | |
Boiler overheat thermostat tripped open prematurely. | Check if overheat thermostat can be manually reset with a cooled boiler. | Replace overheat thermostat. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C982: CLOSE MODE BOILER HI-LIMIT OPEN ITEMS NOT STERILE TURN POWER OFF/ON Boiler overheat thermostat has opened for at least 0.25 seconds & needs to be manually reset. |
Water pump not primed. | Water pump priming instructions. | Prime water pump. |
Boiler temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Replace boiler temperature sensor. | |
Boiler overheat thermostat disconnected at PC board. | Check wire connections at J3 on PC board. | Secure wire connections. | |
Loose or broken wire connections at boiler overheat thermostat. | Check wire connections at boiler overheat thermostat. | Secure wire connections. | |
Boiler overheat thermostat tripped open prematurely. | Check if overheat thermostat can be manually reset with a cooled boiler. | Replace overheat thermostat. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C983: HEATUP MODE BOILER HI-LIMIT OPEN ITEMS NOT STERILE TURN POWER OFF/ON Boiler overheat thermostat has opened for at least 0.25 seconds & needs to be manually reset. |
Water pump not primed. | Water pump priming instructions. | Prime water pump. |
Boiler temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Replace boiler temperature sensor. | |
Boiler overheat thermostat disconnected at PC board. | Check wire connections at J3 on PC board. | Secure wire connections. | |
Loose or broken wire connections at boiler overheat thermostat. | Check wire connections at boiler overheat thermostat. | Secure wire connections. | |
Boiler overheat thermostat tripped open prematurely. | Check if overheat thermostat can be manually reset with a cooled boiler. | Replace overheat thermostat. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C984: STERILIZE MODE BOILER HI-LIMIT OPEN ITEMS NOT STERILE TURN POWER OFF/ON Boiler overheat thermostat has opened for at least 0.25 seconds & needs to be manually reset. |
Water pump not primed. | Water pump priming instructions. | Prime water pump. |
Boiler temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Replace boiler temperature sensor. | |
Boiler overheat thermostat disconnected at PC board. | Check wire connections at J3 on PC board. | Secure wire connections. | |
Loose or broken wire connections at boiler overheat thermostat. | Check wire connections at boiler overheat thermostat. | Secure wire connections. | |
Boiler overheat thermostat tripped open prematurely. | Check if overheat thermostat can be manually reset with a cooled boiler. | Replace overheat thermostat. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C986: OPEN MODE BOILER HI-LIMIT OPEN ITEMS NOT STERILE TURN POWER OFF/ON Boiler overheat thermostat has opened for at least 0.25 seconds & needs to be manually reset. |
Water pump not primed. |
Water pump priming instructions. |
Prime water pump. |
Boiler temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Replace boiler temperature sensor. | |
Boiler overheat thermostat disconnected at PC board. | Check wire connections at J3 on PC board. | Secure wire connections. | |
Loose or broken wire connections at boiler overheat thermostat. | Check wire connections at boiler overheat thermostat. | Secure wire connections. | |
Boiler overheat thermostat tripped open prematurely. | Check if overheat thermostat can be manually reset with a cooled boiler. | Replace overheat thermostat. |
Display / Symptom | Cause | Check | Action |
---|---|---|---|
C987: DRY MODE BOILER HI-LIMIT OPEN ITEMS NOT STERILE TURN POWER OFF/ON Boiler overheat thermostat has opened for at least 0.25 seconds & needs to be manually reset. |
Water pump not primed. | Water pump priming instructions. | Prime water pump. |
Boiler temperature sensor not functioning properly. | Temperature Sensor Resistance Test. | Replace boiler temperature sensor. | |
Boiler overheat thermostat disconnected at PC board. | Check wire connections at J3 on PC board. | Secure wire connections. | |
Loose or broken wire connections at boiler overheat thermostat. | Check wire connections at boiler overheat thermostat. | Secure wire connections. | |
Boiler overheat thermostat tripped open prematurely. | Check if overheat thermostat can be manually reset with a cooled boiler. | Replace overheat thermostat. |