M9/M11 Sterilizer Activating Service Diagnostics Test and Repair
The Service Diagnostics feature allows you to view recent error codes and test the sterilizer’s major components without running a complete cycle. The Service Diagnostics tests should always be done before replacing any major component.
To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
Some procedures require power to be connected with covers removed. Line voltage is present. Use extreme caution to prevent electric shock.
This operation requires power to be connected to the unit with panels removed. Use caution when performing this procedure.
Model |
M9M11 -020 thru -022, -033, -034 |
Serial Number | All |
1.Disconnect sterilizer power cord.
2.Remove RH side panel.
3.Move switch #1 on SW1 block to ON.
4.Reconnect power cord.
5.Press Start button.
Go to Test Selection Screen for next step.
1.Disconnect power cord.
2.Move switch #1 (SW1 block) to OFF.
3.Reconnect power cord.
Model |
M9M11 -040 thru -043 |
Serial Number | All |
1.Disconnect sterilizer power cord.
2.Remove RH side panel.
3.Move switch #1 on SW1 block to ON.
4.Reconnect power cord.
5.Press Start button.
Go to Test Selection Screen for next step.
1.Disconnect power cord.
2.Move switch #1 (SW1 block) to OFF.
3.Reconnect power cord.