625 Table Specifications
|
Model |
625 -001, -003, -005, -006 |
| Serial Number | All |
| 625 Table -001 /-003 /-005 /-006 Specifications | |
|---|---|
|
Patient Weight (maximum): |
650 lbs (295 kg) |
|
Intended Weight: |
Between 30 and 650 pounds (13.6/295kg) |
| Shipping Options and Weights: | |
| Table with Upholstery | 470 lbs (213 kg) |
| Table with Packaging and Skid, no Upholstery | 487 lbs (221 kg) |
| Upholstery with Packaging, shipped separately | 45 lbs (20 kg) |
|
Power Cord Length: |
8 ft. (244 cm) |
|
Electrical Requirements: |
Refer to: Model Identification / Compliance Chart |
|
Foot / Hand Control Voltage: |
14 VAC, SELV (Safety Extra Low Voltage) |
|
Duplex Receptacle(s) (maximum load): |
115 VAC, 5A, 50/60 Hz |
| Fuses: | |
| IEC inlet fuses (2) | 6.3A, 250V, Type-T, 5 x 20 mm |
| F1 and F2, on the main PC board | 10A, 250V, Type-T, 5 x 20 mm |
| F3, on the main PC board | 160mA, 250V, Type-T, 5 x 20 mm |
| F4, on the main PC board | 6.3A, 250V, Type-T, 5 x 20 mm |
|
Models with drawer heaters - Distribution board fuses (2) |
800mA, 250V, Type-T, 5 x 20mm |
| Duty Cycles: | |
|
Motor Run Time: |
Intermittent Operation, 30 seconds ON / 5 minutes OFF |
|
Receptacle(s) and Drawer Heater: |
Continuous Operation |
|
Classifications: |
Class 1, Type B Applied Part |
|
Protection against ingress of fluids: |
IPXO Foot Control only: IPX1 |
Equipment not suitable for use in the presence of a flammable anesthetic mixture.
|
Model |
625 -004 |
| Serial Number | All |
| 625 Table -004 Specifications | |
|---|---|
|
Patient Weight (maximum): |
650 lbs (295 kg) |
|
Intended Weight: |
Between 30 and 650 pounds (13.6/295kg) |
| Shipping Options and Weights: | |
|
Table with Upholstery |
475 lbs (216 kg ) |
| Table with Packaging and Skid, no Upholstery | 502 lbs (228 kg) |
|
Upholstery with Packaging, shipped separately |
45 lbs (20 kg) |
|
Power Cord Length: |
8 ft. (244 cm) |
|
Electrical Requirements: |
Refer to: Model Identification / Compliance Chart |
|
Foot / Hand Control Voltage: |
14 VAC, SELV (Safety Extra Low Voltage) |
|
Simplex Receptacles (maximum load): |
115 VAC, 5A, 50/60 Hz |
| Fuses: | |
| IEC inlet fuses (2) | 6.3A, 250V, Type-T, 5 x 20 mm |
| F1 and F2, on the main PC board | 10A, 250V, Type-T, 5 x 20 mm |
| F3, on the main PC board | 160mA, 250V, Type-T, 5 x 20 mm |
| F4, on the main PC board | 6.3A, 250V, Type-T, 5 x 20 mm |
| Models with drawer heaters - Distribution board fuses (2) | 800mA, 250V, Type-T, 5 x 20mm |
|
Duty Cycles: |
|
| Motor Run Time: | Intermittent Operation , 30 seconds ON / 5 minutes OFF |
| Receptacles and Drawer Heater: | Continuous Operation |
|
Classifications: |
Class 1, Type B Applied Part |
|
Protection against ingress of fluids: |
IPXO Foot Control only: IPX1 |
Equipment not suitable for use in the presence of a flammable anesthetic mixture.