625 Table Top Cover Access Procedures for Back Limit Switch Access
Removal / Installation
1.Remove cladding and table shrouds.
2.Remove three screws securing top cover.
The back limit switches can be accessed without completely removing the top cover. But if removing the cover is necessary, perform Step 3.
3.Partially separate the back lift actuator as shown. Slide top cover off of actuator.
4.Slide top cover over back actuator. Reattach back actuator.
5.Secure top cover with screws removed in Step 2.
6.Re-attach cladding and table shrouds.