640 Tanita Scale Exam Table Scheduled Maintenance Chart

640 Exam Table Scheduled Maintenance
Interval Inspection or Service Service, Adjust , Repair and/or Replace as Required
Daily Cleaning Clean upholstery with diluted bleach solution. Ratio 10:1 water:bleach
Wipe painted metal & plastic surfaces with a clean soft cloth and mild cleaner.
Obvious Damage Visually inspect components for damage that could cause problems during operation or unsafe operation.
Weekly Check accuracy of digital scale Refer to: Digital Scale - Verifying Scale Accuracy
Every two months Check measuring arm alignment Refer to: Measuring Arm Alignment
Semi-Annually Mechanical Operation Check all mechanical functions for proper operation
Verify Table is Level Place a level on table top. Adjust if necessary.
Labels / Decals Replace any missing or illegible labels.
Lubricate Lubricate moving parts with light machine oil.
Hardware All fasteners must be present and fastened securely.
Date of Service _____/____/____  
Location    
Service Technician