222 Exam Table, 223 and 623 Table Specifications

Barrier-Free® Examination Table Specifications Barrier-Free® Examination Table Specifications

222 Exam Table, 223 and 623 Table Specifications

Patient Weight (maximum)

400 lbs (181.4 kgs)

Paper Roll (maximum size):

21 in. long x 3.5 in. diameter (53.3 x 8.9 cm)

Paper Roll Export models

19.7 in. long x 4.7 in. diameter (50 cm x 12 cm)

Range of Motion

See illustration

Dimensions

See illustration

Weight of Table

230 VAC models only

230 V transformer adds 14.5 lbs (6.5 kg)]

w/ upholstery

325 lbs (147.4 kg) 

w/packaging and skid (no upholstery)

375 lbs (192.7 kg)

Weight of Upholstery

w/packaging (shipped separately)

40 lbs (18.1 kg)

Power Cord(s) [115V models have two cords]

8 ft. (243.8 cm) long

Electrical Rating

See Compliance Charts

Foot Control Voltage

4 VAC, SELV (Safety Extra Low Voltage)

Duplex Receptacle maximum load

115 VAC, 5A

Fuses

Refer to: Fuses

Duty Cycle: Actuator motor run time

Intermittent Operation

Domestic: 30 seconds ON / 5 minutes OFF

Export: 25 seconds ON / 7 minutes OFF

Receptacle / DrawerHeater / Uph. Htr

Continuous Operation

Classifications

Class I, Type B Applied Part, Ordinary Equipment, Intermittent Operation

Domestic: 30 seconds ON / 5 minutes OFF

Export: 25 seconds ON / 7 minutes OFF

Protection against ingress of fluids

IPX0 [Foot Control only: IPX1]

Regulatory Compliance

See Regulatory Compliance Charts

Table Speeds by function

Base

14 - 26 seconds

Back (models w/power back)

15 - 20 seconds