222 Exam Table, 223 and 623 Table Base / Back Actuator Test and Repair
This illustration shows the three serviceable components of the actuator. Use the table below to isolate the malfunctioning part.
Problem | Required Action |
---|---|
Motor runs, But makes grinding noise. |
Clean ball screw threads or replace ball screw assembly |
Motor runs, but table does not move. |
Inspect/replace motor coupler. |
Motor does not run. |
Perform Actuator motor test. |
Unplug power cord before removing PC Board cover.
1.Disconnect actuator wires from PC Board. Base Actuator: J2 and J3, Back Actuator: J4 and J5 (223 / 623 only)
2.Set meter to 200 .
3.Place meter probes on actuator wires.
Meter Reading | Status | Required Action |
---|---|---|
2-3 |
|
Motor is good. Perform next check. |
0 |
|
Replace motor. |