631 Chair Troubleshooting Chart
|
Symptom |
Probable Cause |
Check |
Solution |
|---|---|---|---|
|
No functions will operate. No power at chair receptacles. |
Facility supply voltage. |
Power cord connections at wall outlet and Power Entry Module at base of chair. |
Secure power cord connections. |
|
Facility circuit breaker. |
Reset facility circuit breaker if necessary. |
||
|
Power Supply PC Board fuse(s) blown. |
Test F1 and F2 fuses on Power Supply PC Board. |
Replace faulty fuse(s). |
|
|
Loose wire connection. |
Wire connection between Power Entry Module and Power Supply PC Board. |
Secure loose wire connections. |
|
|
Loose Power Toroid wire connections. |
Wire connections at J9 and J10 on Power Supply PC Board. |
Secure loose wire connections. |
|
|
Power Toroid malfunctioning. |
Refer to: Power Toroid |
Replace Power Toroid. |
|
|
No functions will operate. Power at chair receptacles. |
System malfunction. |
Error code on Machine Control PC Board. |
Refer to: Error Codes |
|
Foot / Hand control. |
Try activating functions from both foot and hand controls. |
Replace faulty hand / foot control. |
|
|
Power Toroid malfunctioning. |
Refer to: Power Toroid |
Replace Power Toroid. |
|
Symptom |
Probable Cause |
Check |
Solution |
|---|---|---|---|
|
No power at chair receptacle(s). All other functions work |
Chair circuit breaker. |
If chair circuit breaker is tripped. |
Reset chair circuit breaker located at the base of chair, press to reset. |
|
Loose wire connection. |
Wire connections at J11 on Power Supply PC Board and chair receptacle. |
Secure loose wire connections. |
|
|
Chair Receptacle. |
Voltage at receptacle. |
Replace faulty receptacle. |
|
|
Power Toroid malfunctioning. |
Refer to: Power Toroid |
Replace Power Toroid. |
|
Symptom |
Probable Cause |
Check |
Solution |
|---|---|---|---|
|
When any control button is pressed, the chair sounds a series of three beeps. No functions operate. |
Control Lockout is enabled. |
Refer to: Control Lockout |
Press and hold the Stop and Back Down buttons simultaneously for two seconds. You will hear a single “Beep” to indicate control lockout is enabled / disabled. |
|
When any control button is pressed, nothing happens. No “beeps”, no movement, etc. |
Faulty controller. |
Refer to: Wired Controls |
Replace faulty controller. |
|
Symptom |
Probable Cause |
Check |
Solution |
|---|---|---|---|
|
Tilt, Back, and Foot functions will not operate. Base function does operate. |
System malfunction. |
Error code on Machine Control PC Board. |
Refer to: Error Codes |
|
Power Supply PC Board fuse(s) blown. |
Test F3 and F4 fuses on Power Supply PC Board. |
Replace faulty fuse(s). |
|
|
Loose wire connection. |
Wire connections at J3 on Power Supply PC Board and J13 on Motor Control Hub Board. |
Secure loose wire connections. |
|
|
Wire connections at J16 on Machine Control PC Board and J14 on Motor Control Hub PC Board. |
Secure loose wire connections. |
|
Symptom |
Probable Cause |
Check |
Solution |
|---|---|---|---|
|
No Base Up. |
Active Sensing Technology® activated. |
For obstruction under foot section. |
Remove object from under foot section. |
|
System malfunction. |
Error code on Machine Control PC Board. |
Refer to:Error Codes |
|
|
Malfunctioning Wired Controls. |
Wired Controls. |
Refer to:Wired Controls Malfunctioning |
|
|
Foot Smart Sensor switches. |
Check wire connections at J18 on Machine Control PC Board and at switches. |
Secure wire connections. Refer to:Fuses and Connections |
|
|
Check Green and White wire connections at four pin inline connection next to Motor Control Hub PC Board. |
Secure wire connections. |
||
|
Test switches. Refer to:Smart Sensor Switches |
Replace Switch(es). |
||
|
Base Up limit switch. |
Wire connections at J5 on Power Supply PC Board and at limit switch. |
Secure wire connections. Refer to:Fuses and Connections |
|
|
Test limit switch. Refer to:Base Actuator / Limit Switches |
Replace limit switch. |
||
|
Base Up Overtravel limit switch. |
Wire connections at J5 on Power Supply PC Board and at limit switch. |
Secure wire connections. Refer to:Fuses and Connections |
|
|
Test limit switch. Refer to:Base Actuator / Limit Switches |
Replace limit switch. |
||
|
No Base Up or Base Down. |
Active Sensing Technology® activated. |
For obstruction under foot section. |
Remove object from under foot section. |
|
System malfunction. |
Error code on Machine Control PC Board. |
Refer to:Error Codes |
|
|
Malfunctioning Wired Controls. |
Wired Controls. |
||
|
Foot Smart Sensor switches. |
Check wire connections at J18 on Machine Control PC Board and at switches. |
Secure wire connections. Refer to:Fuses and Connections |
|
|
Check Green and White wire connections at four pin inline connection next to Motor Control Hub PC Board. |
Secure wire connections. |
||
|
Test switches. Refer to:Smart Sensor Switches |
Replace Switch(es). |
||
|
Loose wire connections. |
Wire connections at J5, J6 and J7 on Power Supply PC Board. |
Secure wire connections. Refer to:Fuses and Connections |
|
|
Power PC Board fuse blown. |
Test F5 fuse on Power Supply PC Board. |
Replace faulty fuse. Refer to:Fuses and Connections |
|
|
Base Up Overtravel limit switch. |
Wire connections at J5 on Power Supply PC Board and at limit switch. |
Secure wire connections. Refer to:Fuses and Connections |
|
|
Test limit switch. Refer to:Base Actuator / Limit Switches |
Replace limit switch. |
||
|
Motor runs, but chair does not move. |
Motor coupler broken. |
Motor coupler. Refer to:Base Actuator / Limit Switches |
Replace motor coupler. |
|
Base Up / Down noisy, grinding or squeaking. |
Actuator threads dry. |
Actuator threads. |
Clean and lubricate base actuator threads. Refer to:Base Actuator / Limit Switches |
|
Base Actuator. |
Run chair up and down, if squeaking or grinding continue check base actuator. |
Replace base actuator. |
|
Symptom |
Probable Cause |
Check |
Solution |
|---|---|---|---|
|
No Back Up. |
System malfunction. |
Error code on Machine Control PC Board. |
Refer to:Error Codes |
|
Malfunctioning Wired Controls. |
Wired Controls. |
||
|
Back Up limit switch. |
Wire connections at J12 on Motor Control Hub PC Board and at limit switch. |
Secure wire connections. Refer to:Fuses and Connections |
|
|
Test limit switch. Refer to:Back Actuator / Limit Switches |
Replace limit switch. |
||
|
Back Up Overtravel limit switch. |
Wire connections at J12 on Motor Control Hub PC Board and at limit switch. |
Secure wire connections. Refer to:Fuses and Connections |
|
|
Test limit switch. Refer to:Back Actuator / Limit Switches |
Replace limit switch. |
||
|
No Back Down. |
Active Sensing Technology® activated. |
Position of chair. Refer to:Active Sensing Technology® |
Activate Base Up and Tilt down, then try Back Down again. |
|
System malfunction. |
Error code on Machine Control PC Board. |
Refer to:Error Codes |
|
|
Malfunctioning Wired Controls. |
Wired Controls. |
||
|
Back Down limit switch. |
Wire connections at J12 on Motor Control Hub PC Board and at limit switch. |
Secure wire connections. Refer to:Fuses and Connections |
|
|
Test limit switch. Refer to:Back Actuator / Limit Switches |
Replace limit switch. |
||
|
Back Down Overtravel limit switch. |
Wire connections at J12 on Motor Control Hub PC Board and at limit switch. |
Secure wire connections. Refer to:Fuses and Connections |
|
|
Test limit switch. Refer to:Back Actuator / Limit Switches |
Replace limit switch. |
||
|
No Back Up or Back Down. |
System malfunction. |
Error code on Machine Control PC Board. |
Refer to:Error Codes |
|
Malfunctioning Wired Controls. |
Wired Controls. |
||
|
Loose wire connections. |
Wire connections at J10, J11 and J12 on Motor Control Hub PC Board. |
Secure wire connections. Refer to:Fuses and Connections |
|
|
Motor Control Hub PC Board fuse blown. |
Test F3 fuse on Motor Control Hub PC Board. |
Replace faulty fuse. Refer to:Fuses and Connections |
|
|
Back Up and/or Down Overtravel limit switch. |
Wire connections at J12 on Motor Control Hub PC Board and at limit switch(es). |
Secure wire connections. Refer to:Fuses and Connections |
|
|
Test limit switch(es). Refer to:Back Actuator / Limit Switches |
Replace limit switch(es). |
||
|
Motor runs, but chair does not move. |
Motor coupler broken. |
Motor coupler. Refer to:Back Actuator / Limit Switches |
Replace motor coupler. |
|
Back Up / Down noisy, grinding or squeaking. |
Actuator threads dry. |
Actuator threads. |
Clean and lubricate back actuator threads. Refer to:Back Actuator / Limit Switches |
|
Back Actuator. |
Run back up and down, if squeaking or grinding continue check back actuator. |
Replace back actuator. |
|
Symptom |
Probable Cause |
Check |
Solution |
|---|---|---|---|
|
No Tilt Up. |
Active Sensing Technology® activated. |
For obstruction on the top of column. |
Remove object from the top of the column. |
|
For obstruction under the foot section. |
Remove object from under the foot section. |
||
|
Position of chair. Refer to: Active Sensing Technology® |
Activate Base Up and Back Up, then try Tilt Up again. |
||
|
System malfunction. |
Error code on Machine Control PC Board. |
Refer to:Error Codes |
|
|
Malfunctioning Wired Controls. |
Wired Controls. |
||
|
Seat Smart Sensor switches. |
Wire connections at J18 on Machine Control PC Board and at switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Check two pin inline connection next to Motor Control Hub PC Board. |
Secure wire connections. |
||
|
Test switches. Refer to: Smart Sensor Switches |
Replace switch(es). |
||
|
Foot Smart Sensor switches. |
Wire connections at J18 on Machine Control PC Board and at switches. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Check Green and White wire connections at four pin inline connection next to Motor Control Hub PC Board. |
Secure wire connections. |
||
|
Test switches. Refer to: Smart Sensor Switches |
Replace switch(es). |
||
|
Tilt Up limit switch. |
Wire connections at J8 on Motor Control Hub PC Board and at limit switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Test limit switch. Refer to: Tilt Actuator / Limit Switches |
Replace limit switch. |
||
|
Tilt Up Overtravel limit switch. |
Wire connections at J8 on Motor Control Hub PC Board and at limit switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Test limit switch. Refer to: Smart Sensor Switches |
Replace limit switch. |
||
|
No Tilt Down. |
Foot extension extended. |
If foot extension is extended. |
Move foot extension in or raise foot section. |
|
Active Sensing Technology® activated. |
For obstruction on the top of column. |
Remove object from the top of the column. |
|
|
For obstruction under the foot section. |
Remove object from under the foot section. |
||
|
Position of chair. Refer to: Active Sensing Technology® |
Activate Base Up and Foot Up, then try Tilt Down again. |
||
|
System malfunction. |
Error code on Machine Control PC Board. |
Refer to:Error Codes |
|
|
Malfunctioning Wired Controls. |
Wired Controls. |
||
|
Foot Extension switch. |
Check wire connections at J6 on Machine Control PC Board and at switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Check Black and Red wire connections at four pin inline connection next to Motor Control Hub PC Board. |
Secure wire connections. |
||
|
Test switches. Refer to: Smart Sensor Switches |
Replace switch. |
||
|
Seat Smart Sensor switches. |
Wire connections at J18 on Machine Control PC Board and at switches. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Check two pin inline connection next to Motor Control Hub PC Board. |
Secure wire connections. |
||
|
Test switches. Refer to: Smart Sensor Switches |
Replace switch(es). |
||
|
Foot Smart Sensor switches. |
Wire connections at J18 on Machine Control PC Board and at switches. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Check Green and White wire connections at four pin inline connection next to Motor Control Hub PC Board. |
Secure wire connections. |
||
|
Test switches. Refer to: Smart Sensor Switches |
Replace switch(es). |
||
|
Tilt Down limit switch. |
Wire connections at J8 on Motor Control Hub PC Board and at limit switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Test limit switch. Refer to: Tilt Actuator / Limit Switches |
Replace limit switch. |
||
|
Tilt Down Overtravel limit switch. |
Wire connections at J8 on Motor Control Hub PC Board and at limit switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Test limit switch. Refer to: Tilt Actuator / Limit Switches |
Replace limit switch. |
||
|
No Tilt Up or Tilt Down. |
Active Sensing Technology™ activated. |
For obstruction on the top of column. |
Remove object from the top of the column. |
|
For obstruction under the foot section. |
Remove object from under the foot section. |
||
|
System malfunction. |
Error code on Machine Control PC Board. |
Refer to:Error Codes |
|
|
Malfunctioning Wired Controls. |
Wired Controls. |
||
|
Seat Smart Sensor switches. |
Wire connections at J18 on Machine Control PC Board and at switches. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Check two pin inline connection next to Motor Control Hub PC Board. |
Secure wire connections. |
||
|
Test switches. Refer to: Smart Sensor Switches |
Replace switch(es). |
||
|
Foot Smart Sensor Switches. |
Wire connections at J18 on Machine Control PC Board and at switches. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Check Green and White wire connections at four pin inline connection next to Motor Control Hub PC Board. |
Secure wire connections. |
||
|
Test switches. Refer to: Smart Sensor Switches |
Replace switch(es). |
||
|
Loose wire connections. |
Wire connections at J6, J7 and J8 on Motor Control Hub PC Board. |
Secure loose wire connections. Refer to: Fuses and Connections |
|
|
Motor Control Hub PC Board fuse blown. |
Test F2 fuse on Motor Control Hub PC Board. |
Replace faulty fuse. Refer to: Fuses and Connections |
|
|
Tilt Up and/or Down Overtravel limit switch. |
Wire connections at J8 on Motor Control Hub PC Board and at limit switch(es). |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Test limit switch(es). Refer to: Tilt Actuator / Limit Switches |
Replace limit switch(es). |
||
|
Motor runs, but chair does not move. |
Motor coupler broken. |
Motor coupler. Refer to: Tilt Actuator / Limit Switches |
Replace motor coupler. |
|
Tilt Up / Down noisy, grinding or squeaking. |
Actuator threads dry. |
Actuator threads. |
Clean and lubricate tilt actuator threads. Refer to: Tilt Actuator / Limit Switches |
|
Tilt Actuator. |
Run tilt up and down, if squeaking or grinding continue check tilt actuator. |
Replace tilt actuator. |
|
Symptom |
Probable Cause |
Check |
Solution |
|---|---|---|---|
|
No Foot Up. |
Debris tray not in the stowed position. |
If Debris tray is not completely pushed in. |
Push Debris tray all the way in. |
|
Active Sensing Technology® activated. |
For obstruction under the foot section. |
Remove object from under the foot section. |
|
|
System malfunction. |
Error code on Machine Control PC Board. |
Refer to:Error Codes |
|
|
Malfunctioning Wired Controls. |
Wired Controls. |
||
|
Debris Tray switch. |
Wire connections at J6 on Machine Control PC Board and at switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Test switch. Refer to: Smart Sensor Switches |
Replace Switch. |
||
|
Foot Smart Sensor switches. |
Wire connections at J18 on Machine Control PC Board and at switches. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Check Green and White wire connections at four pin inline connection next to Motor Control Hub PC Board. |
Secure wire connections. |
||
|
Test switches. Refer to: Smart Sensor Switches |
Replace switch. |
||
|
Foot Up limit switch. |
Wire connections at J4 on Motor Control Hub PC Board and at limit switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Test limit switch. Refer to: Foot Actuator / Limit Switches |
Replace limit switch. |
||
|
Foot Up Overtravel limit switch. |
Wire connections at J4 on Motor Control Hub PC Board and at limit switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Test limit switch. Refer to: Foot Actuator / Limit Switches |
Replace limit switch. |
||
|
No Foot Down. |
Foot extension extended. |
If foot extension is extended. |
Move foot extension in. |
|
Debris tray not in the stowed position. |
If debris tray is not completely pushed in. |
Push debris tray all the way in. |
|
|
Active Sensing Technology® activated. |
For obstruction under the foot section. |
Remove object from under the foot section. |
|
|
System malfunction. |
Error code on Machine Control PC Board. |
Refer to:Error Codes |
|
|
Malfunctioning Wired Controls. |
Wired Controls. |
||
|
Foot Extension switch. |
Check wire connections at J6 on Machine Control PC Board and at switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Check Black and Red wire connections at four pin inline connection next to Motor Control Hub PC Board. |
Secure wire connections. |
||
|
Test switches. Refer to: Smart Sensor Switches |
Replace switch. |
||
|
Debris Tray switch. |
Wire connections at J6 on Machine Control PC Board and at switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Test switch. Refer to: Smart Sensor Switches |
Replace switch. |
||
|
Foot Smart Sensor switches. |
Wire connections at J18 on Machine Control PC Board and at switches. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Check Green and White wire connections at four pin inline connection next to Motor Control Hub PC Board. |
Secure wire connections. |
||
|
Test switches. Refer to: Smart Sensor Switches |
Replace switch. |
||
|
Foot Down limit switch. |
Wire connections at J4 on Motor Control Hub PC Board and at limit switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Test limit switch. Refer to: Foot Actuator / Limit Switches |
Replace limit switch. |
||
|
Foot Down Overtravel limit switch. |
Wire connections at J4 on Motor Control Hub PC Board and at limit switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Test limit switch. Refer to: Foot Actuator / Limit Switches |
Replace limit switch. |
||
|
No Foot Up or Foot Down. |
Debris Tray not in the stowed position. |
If debris tray is not completely pushed in. |
Push debris tray all the way in. |
|
Active Sensing Technology® activated. |
For obstruction under the foot section. |
Remove object from under the foot section. |
|
|
System malfunction. |
Error code on Machine Control PC Board. |
Refer to:Error Codes |
|
|
Malfunctioning Wired Controls. |
Wired Controls. |
||
|
Debris Tray switch. |
Wire connections at J6 on Machine Control PC Board and at switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Test switch. Refer to: Smart Sensor Switches |
Replace switch. |
||
|
Foot Smart Sensor switches. |
Wire connections at J18 on Machine Control PC Board and at switches. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Check Green and White wire connections at four pin inline connection next to Motor Control Hub PC Board. |
Secure wire connection. |
||
|
Test switches. Refer to: Smart Sensor Switches |
Replace switch. |
||
|
Loose wire connections. |
Wire connections at J2, J3 and J4 on Motor Control Hub PC Board. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Motor Control Hub PC Board fuse blown. |
Test F1 fuse on Motor Control Hub PC Board. |
Replace faulty fuse. Refer to: Fuses and Connections |
|
|
Foot Up and/or Down Overtravel limit switch. |
Wire connections at J4 on Motor Control Hub PC Board and at limit switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Test limit switch. Refer to: Foot Actuator / Limit Switches |
Replace limit switch. |
||
|
Motor runs, but chair does not move. |
Motor coupler broken. |
Motor coupler. Refer to: Foot Actuator / Limit Switches |
Replace motor coupler. |
|
Foot Up / Down noisy, grinding or squeaking. |
Actuator threads dry. |
Actuator threads. |
Clean and lubricate foot actuator threads. Refer to: Foot Actuator / Limit Switches |
|
Foot Actuator. |
Run foot up and down, if squeaking or grinding continue check foot actuator. |
Replace foot actuator. |
|
Symptom |
Probable Cause |
Check |
Solution |
|---|---|---|---|
|
“Home” function does not lower the chair to a height of 18 inches. |
“Home” button pressed, then released. |
If operator pressed and released “Home” button. |
Press and hold “Home” button. |
|
Foot extension extended. |
If foot extension is extended. |
Move foot extension in. |
|
|
Active Sensing Technology® activated. |
For obstruction on the top of column. |
Remove object from the top of the column. |
|
|
For obstruction under foot section. |
Remove object from under foot section. |
||
|
Position of chair. Refer to: Active Sensing Technology® |
Activate Tilt Up / Down and Back Up, then try Base Down again. |
||
|
System malfunction. |
Error code on Machine Control PC Board. |
Refer to:Error Codes |
|
|
Malfunctioning Wired Controls. |
Wired Controls. |
||
|
Foot Extension switch. |
Wire connections at J6 on Machine Control PC Board and at switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Check Black and Red wire connections at four pin inline connection next to Motor Control Hub PC Board. |
Secure wire connections. |
||
|
Test switch. Refer to: Smart Sensor Switches |
Replace switch. |
||
|
Seat Smart Sensor switches. |
Wire connections at J18 on Machine Control PC Board and at switches. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Check two pin inline connection next to Motor Control Hub PC Board. |
Secure wire connection. |
||
|
Test switches. Refer to: Smart Sensor Switches |
Replace switch(es). |
||
|
Foot Smart Sensor switches. |
Wire connections at J18 on Machine Control PC Board and at switches. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Check Green and White wire connections at four pin inline connection next to Motor Control Hub PC Board. |
Secure wire connection. |
||
|
Test switches. Refer to: Smart Sensor Switches |
Replace switch(es). |
||
|
Base Down limit switch. |
Wire connections at J5 on Power Supply PC Board and at limit switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Test limit switch. Refer to: Base Actuator / Limit Switches |
Replace limit switch. |
||
|
Base Up Overtravel limit switch. |
Wire connections at J5 on Power Supply PC Board and at limit switch. |
Secure wire connections. Refer to: Fuses and Connections |
|
|
Test limit switch. Refer to: Base Actuator / Limit Switches |
Replace limit switch. |