646 Chair Troubleshooting Chart


Symptom |
Probable Cause |
Check |
Correction |
---|---|---|---|
Power light is OFF. |
Facility supply voltage. |
Power cord connections and facility circuit breaker. |
Secure power cord connections. Reset circuit breaker if necessary. |
Primary fuse(s) blown. |
Inspect fuses. |
Replace faulty fuse(s). |
|
Main PC board. |
Wire connections between power cord and main PC board.
|
Secure loose connections. If connections are OK, replace main PC board. |
|
Power light is ON. |
Foot control / touch pad. |
Try activating functions from each touch pad / foot control. |
Refer to: Foot Control / Touch Pads |

Symptom |
Probable Cause |
Check |
Correction |
---|---|---|---|
There is power to the chair, but no power at the chair receptacles. |
Power cord connection. |
Check both power cord connections. |
Secure both power cord connections. |
Receptacle circuit breaker(s). |
Check if receptacle circuit breaker(s) tripped. |
Reset receptacle circuit breaker(s) at rear of chair base. |
|
Loose / damaged wire connections. |
Wire connections between receptacle power cord and table receptacles. |
Secure / repair wire connections. |

Symptom |
Probable Cause |
Check |
Correction |
---|---|---|---|
Foot extension will not lock in place or foot extension will not release. |
Foot extension locking mechanism malfunctioning. |
Perform Locking Mechanism Inspection. Refer to: Foot Extension |
Follow instructions outlined in link. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
No Back Up or Back Down. |
Foot control / touch pad. |
Try activating functions from each touch pad / foot control. |
Refer to: Foot Control / Touch Pads |
Loose / damaged wire connections. |
Check wire connections to back actuator and back limit switches. Check wire connections between main system transformer and main PC board (black and white wires). |
Secure any loose connections. |
|
Back actuator / main PC board. |
Isolate the malfunction. Refer to: Back Actuator / Limit Switches |
Follow test sequence outlined in link. |
|
No Back Up, Back Down OK or No Back Down, Back Up OK. |
Foot control / touch pad. |
Try activating functions from each touch pad / foot control. |
Refer to: Foot Control / Touch Pads |
No Back Up - Back Up limit switch. No Back Down - Back Down limit switch. |
Wire connections to limit switch. |
Replace back limit switch / bracket assembly. Refer to: Back Actuator / Limit Switches |
|
Back actuator / main PC board. |
Isolate the malfunction. Refer to: Back Actuator / Limit Switches |
Follow test sequence outlined in link. |
|
Back drifts down. |
Back actuator / motor coupler. |
Refer to: Back Actuator / Limit Switches |
Replace motor coupler. |
Noisy operation - Grinding, squeaking, etc. |
Back actuator. |
Refer to: Back Actuator / Limit Switches |
Follow instructions outlined in link. |
Back function moves slowly, and/or will not lift patient. |
Patient exceeded 450 lbs weight limit. |
Check if Patient exceeded 450 lbs weight limit. |
Inform staff that max patient weight is 450 lbs. |
Low voltage to table. |
Check supply voltage. Required: 115 VAC +10% |
Connect adequate supply voltage. |
|
Back actuator motor. |
Perform Actuator Motor Test. Refer to: Back Actuator / Limit Switches |
Follow test sequence outlined in link. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
No Tilt Up or Tilt Down. |
Foot control / touch pad. |
Try activating functions from each touch pad / foot control. |
Refer to: Foot Control / Touch Pads |
Loose / damaged wire connections. |
Check wire connections to tilt actuator and tilt down limit switch. |
Secure any loose connections. |
|
Tilt actuator / main PC board. |
Isolate the malfunction. Refer to: Tilt Actuator / Limit Switch / Sensor PC Board |
Follow test sequence outlined in link. |
|
No Tilt Up, Tilt Down OK or No Tilt Down, Tilt Up OK. |
Foot control / touch pad. |
Try activating functions from each touch pad / foot control. |
Refer to: Foot Control / Touch Pads |
No Tilt Down - Crash Avoidance System. |
Isolate the malfunction. Refer to: Tilt Actuator / Limit Switch / Sensor PC Board |
Follow test sequence outlined in link. |
|
No Tilt Down - Tilt Down limit switch. |
Check wire connections to limit switch. |
If connections are OK, perform Limit Switch Test. Refer to: Tilt Actuator / Limit Switch / Sensor PC Board |
|
Tilt actuator / main PC board. |
Isolate the malfunction. Refer to: Tilt Actuator / Limit Switch / Sensor PC Board |
Follow test sequence outlined in link. |
|
Tilt Up / Down OK but beeps in full Up position. |
Sensor PC board. |
Perform Sensor PC Board Check. Refer to: Tilt Actuator / Limit Switch / Sensor PC Board |
Replace Sensor PC Board if necessary. |
Seat drifts down. |
Tilt actuator / motor coupler. |
Replace motor coupler. |
|
Noisy operation - Grinding, squeaking, etc. |
Tilt actuator. |
Follow instructions outlined in link. |
|
Tilt function moves slowly and/or will not lift patient. |
Patient exceeded 450 lbs weight limit. |
Check if Patient exceeded 450 lbs weight limit. |
Inform staff that max patient weight is 450 lbs. |
Low voltage to table. |
Check supply voltage. Required: 115 VAC +10% |
Connect adequate supply voltage. |
|
Tilt actuator motor. |
Perform Actuator Motor Test. Refer to: Tilt Actuator / Limit Switch |
Follow test sequence outlined in link. |


Symptom |
Probable Cause |
Check |
Correction |
---|---|---|---|
Power light is OFF. |
Facility supply voltage. |
Power cord connections and facility circuit breaker. |
Secure power cord connections. Reset circuit breaker if necessary. |
Power light is ON. |
Primary fuse(s) blown. |
Inspect fuses. |
Replace faulty fuse(s). |
Power / QuickChair PC board. |
Wire connections between power cord and Power / Quick Chair PC board. |
Secure loose connections. If connections are OK, replace Power / QuickChair PC board. |
|
Foot control / touch pad. |
Try activating functions from each touch pad / foot control. |
Refer to: Foot Control / Touch Pads |

Symptom |
Probable Cause |
Check |
Correction |
---|---|---|---|
There is power to the chair, but no power at the chair receptacles. |
Power cord connection. |
Check both power cord connections. |
Secure both power cord connections. |
Receptacle circuit breaker(s). |
Check if receptacle circuit breaker(s) tripped. |
Reset receptacle circuit breaker(s) at rear of chair base. |
|
Loose / damaged wire connections. |
Wire connections between receptacle power cord and table receptacles. |
Secure / repair wire connections. |

Symptom |
Probable Cause |
Check |
Correction |
---|---|---|---|
Foot extension will not lock in place or foot extension will not release. |
Foot extension locking mechanism malfunctioning. |
Foot extension locking mechanism. Refer to: Locking Mechanism Inspection |
Replace faulty component(s). |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
No Back Up or Back Down. |
Foot control / touch pad. |
Try activating functions from each touch pad / foot control. |
Refer to: Foot Control / Touch Pads |
Loose / damaged wire connections. |
Check wire connections to Back actuator and Back limit switches. |
Secure any loose connections. |
|
Back actuator / Motor Control PC board. |
Refer to: Back Actuator / Limit Switches |
Follow test sequence outlined in link. |
|
No Back Up, Back Down OK or No Back Down, Back Up OK. |
Foot control / touch pad. |
Try activating functions from each touch pad / foot control. |
Refer to: Foot Control / Touch Pads |
No Back Up - Back Up limit switch. No Back Down - Back Down limit switch. |
Wire connections to limit switch. |
Replace limit switch / bracket assembly. Refer to: Back Actuator / Limit Switches |
|
Back actuator / Motor Control PC board. |
Refer to: Back Actuator / Limit Switches |
Follow test sequence outlined in link. |
|
Back drifts down. |
Back actuator / motor coupler. |
Refer to: Back Actuator / Limit Switches |
Replace motor coupler. |
Noisy operation - Grinding, squeaking, etc. |
Back actuator. |
Refer to: Back Actuator / Limit Switches |
Follow instructions outlined in link. |
Back function moves slowly, and/or will not lift patient. |
Patient exceeded 450 lbs weight limit. |
Check if Patient exceeded 450 lbs weight limit. |
Inform staff that max patient weight is 450 lbs. |
Low voltage to table. |
Check supply voltage. Required: 115 VAC +10% |
Connect adequate supply voltage. |
|
Back actuator motor. |
Perform Actuator Motor Test. Refer to: Back Actuator / Limit Switches |
Follow test sequence outlined in link. |

Symptom | Probable Cause | Check | Correction |
---|---|---|---|
No Tilt Up or Tilt Down. |
Foot control / touch pad. |
Try activating functions from each touch pad / foot control. |
Refer to: Foot Control / Touch Pads |
Loose / damaged wire connections. |
Check wire connections to: tilt actuator, tilt down limit switch. |
Secure any loose connections. |
|
Tilt actuator / Motor Control PC board. |
Follow test sequence outlined in link. |
||
No Tilt Up, Tilt Down OK or No Tilt Down, Tilt Up OK. |
Foot control / touch pad. |
Try activating functions from each touch pad / foot control. |
Refer to: Foot Control / Touch Pads |
No Tilt Down - Crash Avoidance System. |
Follow test sequence outlined in link. |
||
No Tilt Down - Tilt Down limit switch. |
Check wire connections to limit switch. |
If connections are OK, perform Limit Switch Test. Refer to: Tilt Actuator / Limit Switch |
|
Tilt actuator / Motor Control PC board. |
Follow test sequence outlined in link. |
||
Tilt Up / Down OK but beeps in full Up position. |
Tilt Up Sensor PC board. |
Perform Sensor PC Board Check. Refer to: Tilt Actuator / Limit Switch |
Replace Sensor PC Board if necessary. |
Seat drifts down. |
Tilt actuator / motor coupler. |
Replace motor coupler. |
|
Noisy operation - Grinding, squeaking, etc. |
Tilt actuator. |
Follow instructions outlined in link. |
|
Tilt function moves slowly, and/or will not lift patient. |
Patient exceeded 450 lbs weight limit. |
Check if Patient exceeded 450 lbs weight limit. |
Inform staff that max patient weight is 450 lbs. |
Low voltage to table. |
Check supply voltage. Required: 115 VAC +10% |
Connect adequate supply voltage. |
|
Tilt actuator motor. |
Perform Actuator Motor Test. Refer to: Tilt Actuator / Limit Switch |
Follow test sequence outlined in link. |