Procenter Assistant's Unit Troubleshooting

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
With water selector valve City / Bottle toggle switch in either position, no water is present at syringe or water quick connect. | Water selector valve is malfunctioning. | Remove outgoing water line from water selector valve, and check if water is present. |
If no water is present, clean or replace water selector valve. Refer to: Console Mtd. Units, LR Units, or Casework Mtd. Units |
Damaged or disconnected tubing. | Check 1/8” blue tubing, between water selector valve and syringe, and 1/4” blue tubing to water quick connect for damage and proper connections. | Replace damaged tubing. Reconnect tubing if necessary. | |
With water selector valve City / Bottle toggle switch in Bottle position, no water is present at syringe or water quick connect. In City position, water is present. | Water bottle is empty. | - | Fill water bottle with distilled water. |
Water bottle gasket is leaking. | Check for air leaking around bottle cap. | Replace water bottle gasket. | |
Water bottle air regulator is out of adjustment or malfunctioning. | Check air regulator pressure setting. | ||
Water bottle air regulator, in console or L/R arm, is malfunctioning. | Water bottle air regulator output pressure should be 30 PSI. | Replace regulator if necessary. | |
Damaged or disconnected tubing. | Check 1/8” brown tubing, between water bottle air regulator in J-box and water bottle, for damage and proper connections. | Replace damaged tubing. Reconnect tubing if necessary. | |
Check 1/8” blue tubing, between water bottle and water selector valve, for damage and proper connection. | Replace damaged tubing. Reconnect tubing if necessary. | ||
Water selector valve is malfunctioning. | Replace suspect valve with known working valve. |
Replace water selector valve. Refer to: Console Mtd. Units, LR Units, or Casework Mtd. Units |
|
With water selector valve City / Bottle toggle switch in CITY position, no water is present at syringe or water quick connect. In BOTTLE position, water is present. | Water manual shut-off valve in J-box is turned off or malfunctioning. | Check that water manual shut-off valve is ON. | Contact licensed plumber to replace manual shut-off valve. |
Damaged or disconnected tubing. | Check 1/8” blue tubing, between air regulator in J-box and water selector valve, for damage and proper connections. | Replace damaged tubing. Reconnect tubing if necessary. | |
Water selector valve is malfunctioning. | Replace suspect valve with known working valve. |
Replace water selector valve. Refer to: Console Mtd. Units, LR Units, or Casework Mtd. Units |

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
No water present at syringe. Water quick connect works properly. | Pinch valve in assistant’s unit adjusted too tightly. | - | Loosen pinch valve. |
Syringe valve is malfunctioning. | Replace suspect valve with known working valve. | Replace syringe button valve. |

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
No water to quick connect port. Syringe works properly. | Water quick connect valve is malfunctioning. | Replace suspect valve with known working valve. | Replace water quick connect valve. |

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
With ON/OFF lever(s) open, there is insufficient suction at saliva ejector and HVE. | Central vacuum unit is OFF or malfunctioning. | Check central vacuum unit. | Contact service provider for central vacuum unit if necessary. |
Damaged or disconnected tubing. | Check 5/8” grey vacuum tubing, between assistant's unit and central vacuum connection in J-box, for damage and proper connections. | Replace damaged tubing. Reconnect tubing if necessary. | |
Hole plug(s) missing from solids collector | Check solids collector. | Replace hole plug(s) on solids collector. |