647 Chair Troubleshooting Chart

Symptom |
Probable Cause |
Check |
Correction |
---|---|---|---|
Power light is OFF.
|
Facility supply voltage. |
Power cord connections and facility circuit breaker. |
Secure power cord connections. Reset circuit breaker if necessary. |
Primary fuse(s) blown. |
Inspect fuses. |
Replace faulty fuse(s). |
|
Main system transformer
|
Wire connections between power inlet and main PC board. |
Secure loose connections. If connections are OK, test the transformer. Main System Transformer |
|
Models with Rotational Base: Wire connections between rotational base PC board and transformer. |
Secure loose connections. If connections are OK, perform EMI Filter Board Test. Rotational Base Brake System. |
||
Main PC board |
If main system transformer is OK |
Replace main PC board. |
|
Power light is ON.
|
Foot control / touch pad |
Try activating functions from each touch pad / foot control. |
|
System malfunction |
Error code LEDs on main PC board. |

Symptom |
Probable Cause |
Check |
Correction |
---|---|---|---|
No Base Up or Base Down |
Foot control / touch pad |
Try activating functions from each touch pad / foot control. |
|
System malfunction |
Error code LEDs on main PC board. |
||
Loose / damaged wire connections |
Check wire connections to base actuator, base limit switches. Check black and white wire connections between main system transformer and main PC board. | Secure any loose connections. |
|
Base actuator / main PC board |
Follow test sequence. |
||
No Base Up. Base Down OK. OR No Base Down. Base Up OK. |
Foot control / touch pad |
Try activating functions from each touch pad / foot control. |
|
System malfunction |
Error code LEDs on main PC board. |
||
No Base Down Crash Avoidance System |
Refer to: Crash Avoidance System |
Determine appropriate test procedure(s) based on the theory of operation for the Crash Avoidance System. |
|
No Base Up Base Up limit switch. No Base Down Base Down limit switch. |
Wire connections to limit switch. |
If connections are OK, perform Limit Switch Test. Base Actuator / Limit Switch |
|
No Base Down Foot extension switch |
Remove any obstructions under foot section of chair. |
Perform Crash Limit Switch Test. Foot Extension / Crash Limit Switch. |
|
Base actuator / main PC board |
Follow test sequence |
||
Base drifts down. |
Base actuator and gas springs |
With no weight on chair, does base still drift down? |
If YES, replace motor coupler and gas springs. If NO, replace base actuator. Base Actuator / Limit Switches |
Noisy operation, grinding, squeaking, etc. |
Gas springs / scissor mechanism |
- |
Clean / lubricate gas spring pivot joints. Wipe guide bars with a clean, dry cloth. Never lubricate guide bars or scissor mechanism! Gas Springs |
Base actuator |
Follow test sequence |
||
Base function moves slowly, and/or will not lift patient. | Patient exceeded 450 lb weight limit | - | Inform staff that max patient weight is 450 lbs. |
Low voltage to chair |
Check supply voltage. Required: 115 VAC ±10% |
Connect adequate supply voltage. |
|
Base actuator motor |
Perform Actuator Motor Test. Base Actuator / Limit Switches |
Follow test sequence outlined. |
|
Gas springs |
If actuator motor is OK |
Replace gas springs. Gas Springs |
|
Base function moves briefly, then stops and beeps. |
Main PC board needs to be calibrated |
Calibrate main PC board |
|
System malfunction |
Error code LEDs on main PC board. |
||
Base position sensor |
Wire connections to sensor. Perform Output Voltage Test at Sensor. Position Sensors |
Follow test sequence. |
|
Base actuator / main PC board |
Perform Actuator Motor Test. Base Actuator / Limit Switches |
Follow test sequence. |

Symptom |
Probable Cause |
Check |
Correction |
---|---|---|---|
No Back Up or Back Down |
Foot control / touch pad |
Try activating functions from each touch pad / foot control. |
|
Loose / damaged wire connections |
Check wire connections to back actuator and back limit switches. Check wire connections between main system transformer and main PC board, black and white wires. |
Secure any loose connections. |
|
Back actuator / main PC board |
Follow test sequence |
||
No Back Up. Back Down OK. OR No Back Down. Back Up OK. |
Foot control / touch pad | Try activating functions from each touch pad / foot control. | Foot Control / Touch Pads |
No Back Down -Crash Avoidance System |
Determine appropriate test procedure(s) based on the theory of operation for the Crash Avoidance System. |
||
No Back Up - Back Up limit switch No Back Down- Back Down limit switch |
Wire connections to limit switch. |
Replace back limit switch / bracket assembly. Back Actuator / Limit Switches |
|
Back actuator / main PC board |
Follow test sequence |
||
Back drifts down. |
Back actuator / motor coupler |
|
Replace motor coupler. Back Actuator / Limit Switches |
Noisy operation, grinding, squeaking, etc. |
Back actuator |
Follow instructions outlined |
|
Back function moves slowly, and/or will not lift patient. |
Patient exceeded 450 lb weight limit |
- |
Inform staff that max patient weight is 450 lbs. |
Low voltage to chair |
Check supply voltage. Required: 115 VAC ±10% |
Connect adequate supply voltage. |
|
Back actuator motor |
Perform Actuator Motor Test. Back Actuator / Limit Switches |
Follow test sequence outlined |
|
Back function moves briefly, then stops and beeps. |
Main PC board needs to be calibrated |
Calibrate main PC board |
|
Back position sensor |
Wire connections to sensor. Perform Output Voltage Testing at the Sensors. Position Sensors |
Follow test sequence outlined |
|
Back actuator / main PC board |
Perform Actuator Motor Test. Back Actuator / Limit Switches |
Follow test sequence outlined |

Symptom |
Probable Cause |
Check |
Correction |
---|---|---|---|
No Tilt Up or Tilt Down |
Foot control / touch pad |
Try activating functions from each touch pad / foot control. |
|
Loose / damaged wire connections |
Check wire connections to tilt actuator, tilt down limit switch. |
Secure any loose connections. |
|
Tilt actuator / main PC board |
Follow test sequence |
||
No Tilt Up. Tilt Down OK. OR No Tilt Down. Tilt Up OK. |
Foot control / touch pad |
Try activating functions from each touch pad / foot control. |
|
No Tilt Up - Tilt position sensor |
Refer to:Tilt Up Operation for description of proper operation. |
Perform Output Voltage Testing at the Sensors. Position Sensors |
|
No Tilt Down - Tilt Down limit switch |
Wire connections to limit switch. |
If connections are OK, perform Limit Switch Test. Tilt Actuator / Limit Switch |
|
Tilt actuator / main PC board |
Follow test sequence |
||
Seat drifts down. |
Tilt actuator / motor coupler |
|
Replace motor coupler. Tilt Actuator / Limit Switch |
Noisy operation, grinding, squeaking, etc. |
Tilt actuator |
Follow instructions outlined |
|
Tilt function moves slowly, and/or will not lift patient. |
Patient exceeded 450 lb weight limit |
|
Inform staff that max patient weight is 450 lbs. |
Low voltage to chair |
Check supply voltage. Required: 115 VAC ±10% |
Connect adequate supply voltage. |
|
Tilt actuator motor |
Perform Actuator Motor Test. Tilt Actuator / Limit Switch |
Follow test sequence |
|
Tilt function moves briefly, then stops and beeps. |
Main PC board needs to be calibrated |
Calibrate main PC board |
|
Tilt position sensor |
Wire connections to sensor. Perform Output Voltage Testing at the Sensors Position Sensors |
Follow test sequence |
|
Tilt actuator / main PC board |
Perform Actuator Motor Test. Tilt Actuator / Limit Switch |
Follow test sequence |

Symptom |
Probable Cause |
Check |
Correction |
---|---|---|---|
Inoperable functions may include any of the following: Base Down, Back Down, Tilt Up, Tilt Down |
Crash Avoidance System |
Determine appropriate test procedure(s) based on the theory of operation for the Crash Avoidance System. |

Symptom |
Probable Cause |
Check |
Correction |
---|---|---|---|
Positions cannot be programmed. Or When position button is pressed, chair does not move, or moves to wrong position. |
Programming failed. |
|
Reprogram desired position. Position Programming |
Programming process error. |
Error code LEDs on main PC board. |
||
Main PC board needs to be calibrated |
Calibrate main PC board |

Symptom |
Probable Cause |
Check |
Correction |
---|---|---|---|
Rotational brake will not lock. |
Rotational brake mechanism |
Unplug chair power cord. If base rotation locks, perform Rot. Brake Electrical Test. If base does not lock, inspect the mechanical brake components. Rotational Base Brake System. |
Follow test sequence outlined. |
Base wobbles when locked. and/or Grinding noise occurs as the base rotates. |
Debris between upper and lower castings. |
|
Without separating the castings, remove any debris. |
Loose hub screws |
Four screws securing the upper and lower castings. For access instructions Refer to: Separating Upper and Lower Base Castings |
Tighten screws. |
|
Rotation bearings / brake disc |
Inspect needle bearing and brake disc for damage, debris, etc. Separating Upper and Lower Base Castings |
Remove debris or replace base if necessary. |
|
Rotational brake will not unlock. |
Brake pedal switch |
Check both brake pedals. |
If either pedal works properly, check connections to faulty pedal switch. Replace switch if necessary. Rotational Base Brake System. |
Electro-magnet |
Perform Electro-magnet Test. Rotational Base Brake System |
Follow test sequence |

Problem |
Symptom |
Probable Cause |
Check |
Correction |
---|---|---|---|---|
No power at chair receptacles.
|
There is power to the chair, but no power at the chair receptacles.
|
Loose / damaged wire connections |
Wire connections between power inlet and chair receptacles |
Correction Secure / repair wire connections. |
Isolation transformer / receptacles |
Perform Output Voltage Test. Chair Receptacles Isolation Transformer |
Follow test sequence. |

Problem |
Symptom |
Probable Cause |
Check |
Correction |
---|---|---|---|---|
Foot extension malfunctioning. |
Foot extension will not lock in place. Or Foot extension will not release. |
Foot extension locking mechanism malfunctioning. |
Perform Locking Mechanism Inspection. Foot Extension. |
Follow instructions. |