255 Light Troubleshooting Chart

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
Light does not function when light is ON. | Facility circuit breaker is tripped. | Check facility circuit breaker. | Reset circuit breaker. |
Line fuse(s) blown. | Inspect fuses. |
Replace fuse(s). Ceiling Mounted Lights, Wall Mtd. Lights, Mobile Lights |
|
Fuse holder malfunctioning. | Perform continuity check on fuse holder. | Replace fuse holder. | |
Switch membrane malfunctioning. | Perform continuity check on switch. | Replace switch membrane. | |
Power supply malfunctioning. | Check connections and voltage. | If proper voltage is not present, replace power supply. | |
Driver board malfunctioning. | Check connections and voltage. | If proper voltage is not present, replace driver board. | |
LED light module malfunctioning. | Check connections and voltage. | If proper voltage is not present, replace LED light module. | |
Damaged / disconnected wiring in the arm assemblies. | Perform continuity check on wiring from transformer to lighthead. |
Repair / replace any damaged or disconnected wiring. Refer to: Wiring Diagrams: Ceiling Mounted Lights, Wall Mtd. Lights, Mobile Lights |
|
Light does not function, or functions intermittently when lighthead or arm assemblies are moved. | Damaged / disconnected wiring in the arm assemblies. | Perform continuity check on wiring from transformer to lighthead. | Repair / replace any damaged or disconnected wiring. Refer to: Wiring Diagrams: Ceiling Mounted Lights, Wall Mtd. Lights, Mobile Lights |

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
Cross tube (at ball pivot) can be rotated more than 580° in one direction. | Ball pivot stop is broken. | Inspect ball pivot stop. | Replace ball pivot stop. |
Mobile light cross tube can be rotated more than 540° in one direction. | Ball pivot stop is broken. | Inspect ball pivot stop stop. | Replace ball pivot stop. |
Lighthead can be rotated more than 180° in one direction. | Lighthead pivot stop is broken. | Inspect lighthead pivot. | Replace lighthead pivot stop. |
Ceiling Mtd Lights only: Down tube can be rotated more than 580° in one direction. | Main pivot stop is broken. | Inspect main pivot stop. | Replace main pivot stop. |
Wall Mtd Lights only: Fixed arm can be rotated more than 170° (collides with wall). | Stop screw is missing / broken. | Inspect fixed arm stop screw. | Replace stop screw. |

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
Cross tube rotates (at ball pivot joint) too tightly, or drifts when released. | Cross tube counterbalance out of adjustment. | Release tension from ball pivot joint and observe to see if the cross tube counterbalances correctly. | Release tension from ball pivot joint and observe to see if the cross tube counterbalances correctly. |
Lighthead does not rotate freely (at cross tube), or drifts when released. | Lighthead tension out of adjustment. | Lighthead tension out of adjustment. | Lighthead adjustment |
Ceiling Mtd Lights only: Down tube does not rotate, or drifts when released. | Improper installation of ceiling plate. Plate deflecting too much, or not level. | Improper installation of ceiling plate. Plate deflecting too much, or not level. | If ceiling plate is not level, readjust as necessary. If there is excessive flexing, support ceiling structure as necessary. Refer to: Installation Guide. |
If ceiling plate is not level, readjust as necessary. If there is excessive flexing, support ceiling structure as necessary Refer to: Installation Guide. | Improper installation of wall mount plate, or fixed arm not level. | Improper installation of wall mount plate, or fixed arm not level. | Adjust fixed arm leveling screws as necessary Refer to: Installation Guide. |

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
Base moves when casters are set. | Base casters are worn or malfunctioning. | Inspect base casters. | Replace base caster(s) as necessary. |