355 Light Troubleshooting Chart

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
Light does not illuminate. ON / STANDBY switch is ON. | Bulb is burned out. | Inspect bulb. | Replace bulb. |
Facility circuit breaker is tripped. | Check facility circuit breaker. | Reset circuit breaker. | |
Line fuse(s) blown. | Inspect fuses. | Replace fuse(s). | |
Fuse holder malfunctioning. | Perform continuity check on fuse holder. | Replace fuse holder. | |
Bulb socket is corroded resulting in bad connection. | Inspect / clean bulb socket. | Replace bulb socket. | |
ON / STANDBY switch malfunctioning. | Perform continuity check on switch. | Replace ON / STANDBY switch. | |
Transformer malfunctioning. | With the ON / STANDBY switch ON, check for 25-28 VAC across output leads of transformer. | If 25-28 VAC is not present, Replace transformer. | |
Damaged / disconnected wiring in the arm assemblies. | Perform continuity check on wiring from transformer to lighthead. | Repair / replace any damaged or disconnected wiring. Refer to: Ceiling Mounted, Wall Mounted, or Mobile Wiring Diagram | |
Mobile Lights only: AC inlet is malfunctioning. | Perform continuity check on AC inlet. | Replace AC inlet. | |
Lighthead will not illuminate, or flashed intermittently when lighthead or arm assemblies are moved. | Damaged / disconnected wiring in the arm assemblies. | Perform continuity check on wiring from transformer to lighthead. | Repair / replace any damaged or disconnected wiring. Refer to: Ceiling Mounted, Wall Mounted, or Mobile Wiring Diagram |

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
Cross tube rotates at ball pivot joint too stiffly or drifts when released. | Cross tube counterbalance out of adjustment. | Release tension from ball pivot joint and observe to see if the cross tube counterbalances correctly. |
Perform Cross tube counterbalance adjustment. Then, perform Ball joint tension adjustment. |
Ball pivot joint tension out of adjustment. | Check for looseness / stiffness in joint. | ||
Lamp tube does not rotate freely at cross tube or drifts when released. | Lamp tube brake tension out of adjustment. | Check for looseness / stiffness in joint. | Lamp tube brake adjustment. |
Lighthead does not rotate freely at lamp tube or drifts when released. | Lighthead brake tension out of adjustment. | Check for looseness / stiffness in joint. | Lighthead brake adjustment. |
Ceiling Mtd Lights only: Down tube does not rotate, or drifts when released. | Improper installation of ceiling plate. Plate deflecting too much, or not level. | Inspect ceiling plate. Check for level and excessive flexing. | If ceiling plate is not level, readjust as necessary. If there is excessive flexing, support ceiling structure as necessary. Refer to: 120V or 230V Ceiling Mounted Light Installation Guide |
Wall Mtd Lights only: Fixed arm does not rotate freely, or drifts when released. | Improper installation of wall mount plate or fixed arm not level. | Inspect wall mount plate. Place a level on fixed arm. | Adjust fixed arm leveling screws as necessary. Refer to: Wood Framed or Steel Framed Wall Mounted Light Installation Guide |

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
Cross tube can be rotated more than 580° in one direction. | Ball pivot stop is broken. | Inspect ball pivot stop. | Replace ball pivot stop. |
Lamp tube can be rotated more than 540° in one direction. | Lamp tube pivot stop is broken. | Inspect lamp tube pivot stop. | Replace lamp tube pivot stop. |
Lighthead can be rotated more than 180° in one direction. | Lighthead pivot bearing stop is broken. | Inspect lighthead pivot bearing. | Replace lighthead pivot bearing. |
Ceiling Mtd Lights only: Down tube can be rotated more than 580° in one direction. | Main pivot stop is broken. | Inspect pivot stop. | Replace main pivot stop. |
Wall Mtd Lights only: Fixed arm can be rotated more than 170° (collides with wall). | Stop screw missing / broken. | Inspect fixed arm stop screw. | Replace fixed arm stop screw. |

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
Lighthead assembly illuminates but not at the proper intensity. | Bulb malfunctioning. | - | Replace bulb. |
Wrong input tap is being used on transformer or facility voltage has changed since installation. | Check facility input voltage and transformer input tap being used. | Select proper transformer input tap based on facility voltage. | |
Transformer malfunctioning. | With the ON / STANDBY switch ON, check for 25-28 VAC across output leads of transformer. | If 25-28 VAC is not present, Replace transformer. | |
Bulb socket corroded resulting in bad connection or high resistance. | Inspect / clean bulb socket. | Replace bulb socket. | |
IR filter is clouded or dirty. | Inspect IR filter. | Clean IR filter. | |
Faceplate is clouded or dirty. | Inspect faceplate. | Clean faceplate. | |
Components of lighthead damaged to the point that lighthead is no longer in tolerance. | - | Return lighthead to Midmark® for repair. Call 1.800.Midmark, follow the prompts for Technical Service. |

Symptom | Probable Cause | Check | Solution |
---|---|---|---|
Base moves when caster brakes are set. | Caster brakes are worn or malfunctioning. | Inspect caster brakes. | Replace brake caster(s) as necessary. |