Midmark M9® -05x / M11® -05x Sterilizer Troubleshooting 07 Control Error Codes
To prevent risk of shock always disconnect power before removing covers or performing any service procedure.
Some procedures require power to be connected with covers removed. Line voltage is present. Use extreme caution to prevent electric shock.

Error Code Display Screen | Cause | Action |
---|---|---|
0741-0A[m]8*: Power lost during [m] Mode
Modes 2, 3, 4: Instruments not sterile and need to be reprocessed. Modes 5, 6, 7: Risk of recontaminated instruments. Any other Mode: [not applicable] |
Power was lost during a sterilization cycle (e.g., could be due to an electrical storm, brownout). |
Secure power cord connections at wall outlet and sterilizer receptacle. Unplug unit for 60 seconds; then plug into a dedicated, properly rated circuit. |
Loose/damaged wire connections. |
Secure/repair AC terminals and connections at J28 on main PC board. |
|
Power cord / receptacle damaged. |
Replace power cord / receptacle. |
|
Error code will not reset. |
Unplug unit for 60 seconds. Replace main PC board. |
|
*[m] Modes: 0 = Power Up, 1 = Ready/Idle/Select Cycle, 2 = Filling, 3 = Heating, 4 = Sterilizing, 5 = Venting, 6 = Door Opening, 7 = Drying |

Error Code Display Screen | Cause | Action |
---|---|---|
0741-070C: Error encountered during Power Up |
Error encountered when initializing during power up.
|
Secure power cord connections at wall outlet and at sterilizer receptacle. Unplug unit for 60 seconds; then plug into a dedicated, properly rated circuit.
|
Check supply voltage and contact an electrician to modify voltage, if needed. |
||
Loose/damaged wire connections. |
Secure/repair AC terminals and connections at J28 on main PC board. |
|
Power cord / receptacle damaged. |
Replace power cord / receptacle. |
|
Error code will not reset. |
Unplug unit for 60 seconds. Replace main PC board. |

Error Code Display Screen | Cause | Action |
---|---|---|
0743-0328: Chamber water low during Filling Mode
Instruments not sterile and need to be reprocessed. |
The chamber water level sensor did not detect that there is enough water in the chamber to complete a cycle. |
Check water level in reservoir and add water, if needed, until water reaches appropriate level. Press STOP button; then initiate a new cycle. |
Chamber filter clogged. |
Clean/replace filter. Clean with Speed-Clean and brush. |
|
Water Level sensor malfunctioning/dirty. |
Clean/dry water level sensor. Clean with Speed-Clean and Scotch-Brite pad. |
|
Secure/repair loose or damaged wire connections. |
||
Perform Water Level Sensor Test. Replace water level sensor. |
||
Air valve malfunctioning/dirty. |
With door open and door switch engaged, initiate fill cycle to verify if filling. • If filling, air valve malfunctioning/dirty. Perform Air Valve Test. Clean/replace air valve. • If not filling, refer to Fill valve malfunctioning/dirty (below). |
|
Fill valve malfunctioning/dirty. |
Inspect fill valve / related tubing for restrictions. Clean/replace fill valve. |
|
With the sterilizer in Filling mode and door switch engaged, perform Fill Valve Continuity Test. Clean/replace fill valve. |
||
PC board malfunctioning. |
Verify voltage to Fill Valve. Perform Fill Valve Supply Voltage Test. Replace Main PC board. |
|
Chamber is dirty or corroded. |
Clean bottom of the chamber with Speed-Clean and distilled water or water that meets the referenced water purity specifications in User Guide. |
|
Restricted condensing coil. |
Check for blockage in coil and assure weep hole is present and not obstructed. Clear blockage / replace condensing coil. |
|
Blockage or malfunction in VistaCool System (if used). |
Check for debris or blockage in the tubing/fittings to the VistaCool System. Clean/replace as required. |
|
Check operation of In-Iine Thermal Sensor. Replace Thermal Regulator Wax Motor or Thermal Sensor. |
||
Check position of tank for proper installation. Install VistaCool as recommended. Refer to: VistaCool Installation Guide and Owners manual. |

Error Code Display Screen | Cause | Action |
---|---|---|
0744-0824: Door switch opened during Filling Mode
Instruments not sterile and need to be reprocessed.
|
Door detected as open during Filling mode. |
Check with operator to determine if door was opened during fill mode. Close door. Cycle will continue where it left off. |
Door/dam gasket(s). |
Inspect door/dam gaskets for proper installation and any signs of sticking. Install door/dam gaskets properly. |
|
If new gasket, check if wire ring in gasket is preventing gasket from fully being seated. Temporarily remove wire ring. Re-install wire ring after gasket is fully seated. |
||
Door switch lever activation obstructed. |
Check for full movement of door switch lever. Remove any obstruction. |
|
Tray rack / tray plate not installed properly. |
Install tray rack / tray plate properly. The 45 degree angled end of tray plate must be positioned upward and to the back of the chamber. |
|
Door switch / PC board malfunctioning. |
Test switch with the door OPEN and CLOSED. |
|
Perform Door Switch Continuity Test. With door closed, disconnect door switch connectors J13 and J24 of main PC board. Place meter probes on door switch wires. If Continuity is not present, replace door switch. Refer to: Door Switch Continuity Test |
||
Check supply voltage from PC board (line voltage 115 or 230 VAC) for door switch and 3.3 VDC for door dry position switch. Refer to: Door Switch Supply Voltage Test Replace Main PC board. |
||
Door motor system / latch “hanging up”. |
With the door closed, run the Door Open Test. Watch for any mechanical issues that may prevent full rotation of door motor cam. Refer to:Component Check Test Replace door motor. |

Error Code Display Screen | Cause | Action |
---|---|---|
0744-0838: Door switch opened during Heating Mode
Instruments not sterile and need to be reprocessed. |
Door detected as open during Heat Up mode. |
Check with operator to determine if door was opened during Heat Up mode. Instruments not sterile and need to be reprocessed. |
Door/Dam gasket(s). |
Inspect door/dam gaskets for proper installation and any signs of sticking. Install door/dam gaskets properly. |
|
If new gasket, check if wire ring in gasket is preventing gasket from fully being seated. Temporarily remove wire ring. Re-install wire ring after gasket is fully seated. |
||
Door switch lever activation obstructed. |
Check for full movement of door switch lever. Remove any obstruction. |
|
Tray rack / tray plate not installed properly. |
Install tray rack / tray plate properly. The 45 degree angled end of tray plate must be positioned upward and to the back of the chamber. |
|
Door pins / latch binding. |
Check for dirty door pins/latch mechanism. Clean door pins / latch mechanism. Clean with Spray on Synthetic Dry Protectant (Aerosol). |
|
Inspect latch mechanism for wear/damage. Replace worn/damaged components. |
||
Door motor system / latch “hanging up”. |
Remove the door cover and RH side panel. With the door closed, run the Door Open Test. Watch for any mechanical issues that may prevent the door from opening. Refer to:Component Check Test Replace worn/damaged components. |
|
Remove the door cover and RH side panel. With the door closed, run the Door Open Test. Watch for any mechanical issues that may prevent the door from opening. Refer to:Component Check Test Replace door motor. |
||
Door hinge binding. |
Check for binding or buildup of debris in the door hinges. Clean debris from hinges. |
|
Door switch / PC board malfunctioning. |
Test switch with the door OPEN and CLOSED. |
|
Perform Door Switch Continuity Test. With door closed, disconnect door switch connectors J13 and J24 of main PC board. Place meter probes on door switch wires. If Continuity is not present, replace door switch. Refer to: Door Switch Continuity Test |
||
Check supply voltage from PC board (line voltage 115 or 230 VAC). Refer to: Door Switch Supply Voltage Test Replace Main PC board. |

Error Code Display Screen | Cause | Action |
---|---|---|
0744-084C: Door switch opened during Sterilizing Mode
Instruments not sterile and need to be reprocessed. |
Door detected as open during Sterilization mode. |
Check with operator to determine if door was opened during Sterilization mode. Instruments not sterile and need to be reprocessed. |
Door switch / PC board malfunctioning. |
Test switch with the door OPEN and CLOSED. |
|
Perform Door Switch Continuity Test. With door closed, disconnect door switch of main PC board. Place meter probes on door switch wires. If Continuity is not present, replace door switch. Refer to:Door Switch Continuity Test |
||
Check supply voltage from PC board (line voltage 115 or 230 VAC). Refer to: Door Switch Supply Voltage Test Replace Main PC board. |

Error Code Display Screen | Cause | Action |
---|---|---|
0744-0858: Door switch opened during Venting Mode
Risk of recontaminated instruments. |
Door detected as open during Venting mode. |
Check with operator to determine if door was opened during Venting mode. Instruments not sterile and need to be reprocessed. |
Door switch / PC board malfunctioning. |
Test switch with the door OPEN and CLOSED. |
|
Perform Door Switch Continuity Test. With door closed, disconnect door switch connectors J13 and J24 of main PC board. Place meter probes on door switch wires. If Continuity is not present, replace door switch. Refer to:Door Switch Continuity Test |
||
Check supply voltage from PC board (line voltage 115 or 230 VAC). Refer to: Door Switch Supply Voltage Test Replace Main PC board. |

Error Code Display Screen | Cause | Action |
---|---|---|
0744-0968: Door switch not detecting door opening
Risk of recontaminated instruments.
|
Door detected as open during Door Open mode. |
Check with operator to determine if door was opened during Door Open mode. Check the door switch. Risk of recontaminated instruments. |
Door/Dam gasket(s). |
Inspect door/dam gaskets for proper installation and any signs of sticking. Install door/dam gaskets properly. |
|
If new gasket, check if wire ring in gasket is preventing gasket from fully being seated. Temporarily remove wire ring. Re-install wire ring after gasket is fully seated. |
||
Trays not pushed in properly. |
Remove obstructions that prevent the trays from sliding in completely. |
|
Door pins / latch binding. |
Check for dirty door pins/latch mechanism. Clean door pins / latch mechanism with Spray on Synthetic Dry Protectant (Aerosol). |
|
Tighten latch mechanism. |
||
Inspect latch mechanism for wear/damage. Replace worn/damaged components. |
||
Door motor system / latch “hanging up”. |
Remove the door cover and RH side panel. With the door closed, run the Door Open Test. Watch for any mechanical issues that may prevent the door from opening. Refer to:Component Check Test Replace worn/damaged components. |
|
Door motor defective. |
Remove the door cover and RH side panel. With the door closed, run the Door Open Test. Watch for any mechanical issues that may prevent the door from opening. Refer to:Component Check Test Replace door motor. |
|
Door hinge binding. |
Check for binding or buildup of debris in the door hinges. Clean debris from hinges. |
|
If debris cannot be cleaned from hinges to remove binding, send unit to Midmark for factory Repair and Return. |
||
Door switch / PC board malfunctioning. |
Test switch with the door OPEN and CLOSED. |
|
Perform Door Switch Continuity Test. With door closed, disconnect door switch connectors J13 and J24 of main PC board. Place meter probes on door switch wires. If Continuity is not present, replace door switch. Refer to: Door Switch Continuity Test |
||
Check supply voltage from PC board (line voltage 115 or 230 VAC). Refer to: Door Switch Supply Voltage Test Replace Main PC board. |
||
Door switch lever activation obstructed. |
Check for full movement of door switch lever. Remove any obstruction. |
|
Residual pressure in chamber. |
Check for chamber pressure when door motor is energized. Clear any restriction that could prevent chamber pressure from reaching the pressure transducer on PC board. |
|
Check the load size. Refer to User Guide. Verify that nothing is touching or crowding the temperature sensor. Reduce load size. |
||
If pressure is above 0.72 psi (5 kPa), check for blockage in the pressure sensor tubing between the PC board and the chamber. Clear blockage / replace pressure transducer tubing. |
||
If chamber filter is clogged, clean/replace filter. Clean with Speed-Clean and brush. |

Error Code Display Screen | Cause | Action |
---|---|---|
0746-06[m]C*: Chamber temperature high during [m] Mode
Modes 2, 3, 4: Instruments not sterile and need to be reprocessed. Modes 5, 6, 7: Risk of recontaminated instruments. Any other Mode: [not applicable] |
||
Office running back-to-back cycles. |
Check if operator is running consecutive cycles without completing the DRY cycle. Complete full cycle. |
|
If the unit skips the Filling mode... |
||
Tray rack / tray plate not installed properly. |
Install tray rack / tray plate properly. The 45 degree angled end of tray plate must be positioned upward and to the back of the chamber. |
|
(M11 only) Mesh filter not in place. |
If filter is not in place, re-install filter. | |
Water level sensor malfunctioning. Dirty/foreign object touching sensor. |
Inspect water level sensor for foreign objects. burs, instruments, etc. Remove foreign objects. Clean/dry water level sensor. Clean with Speed-Clean and Scotch-Brite pad. Perform Water Level Sensor Test. Replace water level sensor. |
|
If the unit completes the Filling mode… |
||
Sterilizer is unlevel. |
Verify support surface is level. Remove tray/trays. Verify water on all sides of chamber is equal. Verify that nothing is touching or crowding the temperature sensor (e.g., trays, pouches, cassettes). |
|
Temperature sensor malfunctioning. |
Check for residue buildup on temperature sensor probe. Clean with Speed-Clean, distilled water, and abrasive pad. |
|
Verify that nothing is touching or crowding the temperature sensor (e.g., trays, pouches, cassettes). |
||
Perform Temperature Sensor Resistance Test. Allow sensor to cool to room temperature before testing. Acceptable range: 1.07 to 1.1K Ω Refer to: Temperature Sensor Resistance Test Replace temperature sensor. |
||
Perform Temperature Sensor Supply Voltage Test. Disconnect sensor harness from J6 on PC board. On pins 1 and 2, acceptable range is 4.0 to 6.0 VDC. Refer to: Temperature Sensor Supply Voltage Test Replace main PC board. |
||
Heating element malfunctioning. |
Inspect heating element. Heating element should lay flat in the bottom of the chamber and be free of any flaking/pitting. Perform Heating Element Resistance Test (should read 9 to 11 Ω for 115V model and 39 to 42 Ω for 230V model). Replace heating element. |
|
PC board malfunctioning. |
Perform Heating Element Supply Voltage Test. With harness attached to leads, should have line voltage in test mode. Perform Heater Relay PC Board Test (Control and Power) Control Test: 3.3 VDC - ON, approx. 0VDC - OFF Power Test: 115/230VAC - ON/approx. 0VAC - OFF Perform Main PC Board Pressure Transducer Voltage Test. Acceptable range: 4.0 to 6.0 VDC |
|
Chamber filter clogged. |
Clean/replace filter. Clean with Speed-Clean and brush. |
|
Pressure tubing at incorrect angle. |
Check tubing for restrictions and ensure tubing is angled down and away from the PC Board. Adjust pressure tubing to correct angle. |
|
Door spring missing / damaged. |
If the door motor releases the latch, but the door remains closed, Inspect door spring for damage. Verify if door is at the vented position. Push in on door with out lifting handle. Replace door spring |
|
Facility voltage too high/low. |
Test facility supply voltage. Voltage must be: • 115V models: 104 to 127 VAC 50/60 Hz • 230V models: 207 to 253 VAC 50/60 Hz Contact an electrician to modify voltage. |
|
Pressure leaks. |
Check for loose/worn door and dam gaskets or water leaking around door. Verify door and dam gaskets are installed properly. Replace gasket kit 1x year. Clean/replace gaskets. |
|
Leaking or defective vent valve: Check for water leaking from condensing coil spout. Perform Vent Valve Test. Clean/replace vent valve. |
||
Leaking or defective fill valve: Check for water leaking back into reservoir through the fill line. Look for bubbles coming from bottom of reservoir. Perform Fill Valve Test. Clean/replace fill valve. |
||
Leaking or defective air valve: Check for excessive steam coming through condensing coil spout. During the HEATING and VENTING modes, it is normal for steam to be exhausted through the spout. Clean/replace air valve. |
||
Leaking or defective Pressure Relief Valve: Check for water/steam leakage from beneath the rear of the sterilizer. Perform Pressure Relief Valve Test. Replace pressure relief valve. |
||
Pressure transducer tubing leaking: Check for steam leakage at pressure transducer tubing connections. Secure pressure transducer tubing connections with high temperature cable ties. |
||
Check all plumbing fitting connections for leakage. Tighten or replace fittings. |
||
*[m] Modes: 0 = Power Up, 1 = Ready/Idle/Select Cycle, 2 = Filling, 3 = Heating, 4 = Sterilizing, 5 = Venting, 6 = Door Opening, 7 = Drying |

Error Code Display Screen | Cause | Action |
---|---|---|
0746-07[m]C*: Chamber temperature sensor malfunction during [m] Mode
Instruments not sterile and need to be reprocessed. |
Unit was brought in from cold environment. |
Allow unit to warm to room temperature. Unplug unit for 60 seconds, then plug unit into a dedicated, properly rated circuit. Allow unit to reach room temperature. |
Chamber temperature reported as less than 32 °F or greater than 392 °F.
|
Check for residue buildup on temperature sensor probe. Clean sensor probe with Speed-Clean, distilled water, and abrasive pad. |
|
Verify that nothing is touching or crowding the temperature sensor (e.g., trays, pouches, cassettes). |
||
Check temperature sensor harness connection at J6 on PC board. Secure/repair loose or damaged connections. |
||
Perform Temperature Sensor Resistance Test. Allow sensor to cool to room temperature before testing. Acceptable range: 1.07 to 1.1K Ω Refer to:Temperature Sensor Resistance Test Replace temperature sensor. |
||
Perform Temperature Sensor Supply Voltage Test. Disconnect sensor harness from J12 on PC board. On pins 1 and 2, acceptable range is 4.0 to 6.0 VDC. Refer to: Temperature Sensor Supply Voltage Test Replace main PC board. |
||
Sterilizer is unlevel. |
Verify support surface is level. Remove tray/trays. Verify water on all sides of chamber is equal. |
|
*[m] Modes: 0 = Power Up, 1 = Ready/Idle/Select Cycle, 2 = Filling, 3 = Heating, 4 = Sterilizing, 5 = Venting, 6 = Door Opening, 7 = Drying |

Error Code Display Screen | Cause | Action |
---|---|---|
0746-033C: Chamber temperature low during Heating Mode
Instruments not sterile and need to be reprocessed. |
Maintenance not performed. |
Verify weekly and monthly maintenance has been performed by customer. Refer to User Guide. Perform monthly maintenance. |
Sterilizer is unlevel. |
Verify support surface is level. Remove tray/trays. Verify water on all sides of chamber is equal. |
|
Heating element malfunctioning. |
Inspect heating element. Heating element should lay flat in the bottom of the chamber and be free of any flaking/pitting. Perform Heating Element Resistance Test (should read 9 to 11 Ω for 115V model and 34 to 42 Ω for 230V model). Replace heating element. |
|
Temperature sensor malfunctioning. Sterilizing mode not reached within 74 minutes of starting Heating mode. |
Check for residue buildup on temperature sensor probe. Clean with Speed-Clean, distilled water, and abrasive pad. |
|
Verify that nothing is touching or crowding the temperature sensor (e.g., trays, pouches, cassettes). |
||
Perform Temperature Sensor Resistance Test. Allow sensor to cool to room temperature before testing. Acceptable range: 1.07 to 1.1K Ω Refer to: Temperature Sensor Resistance Test Replace temperature sensor. |
||
Perform Temperature Sensor Supply Voltage Test. Disconnect sensor harness from J6 on PC board. On pins 1 and 2, acceptable range is 4.0 to 6.0 VDC. Refer to: Temperature Sensor Supply Voltage Test Replace main PC board. |
||
Pressure leaks. |
Check for loose/worn door and dam gaskets or water leaking around door. Verify door and dam gaskets are installed properly. Replace gasket kit 1x year. Clean/replace gaskets. |
|
Check for voltage flow through harness. Refer to:Fill / Vent Valve Supply Voltage Test Replace harness. |
||
Leaking or defective vent valve: Check for water leaking from condensing coil spout. Perform Vent Valve Test. Clean/replace vent valve. |
||
Leaking or defective fill valve: Check for water leaking back into reservoir through the fill line. Look for bubbles coming from bottom of reservoir. Perform Fill Valve Test. Clean/replace fill valve. |
||
Leaking or defective air valve: Check for excessive steam coming through condensing coil spout. During the HEATING and VENTING modes, it is normal for steam to be exhausted through the spout. Clean/replace air valve. |
||
Leaking or defective Pressure Relief Valve: Check for water/steam leakage from beneath the rear of the sterilizer. Perform Pressure Relief Valve Test. Replace pressure relief valve. |
||
Pressure transducer tubing leaking: Check for steam leakage at pressure transducer tubing connections. Secure pressure transducer tubing connections with high temperature cable ties. |
||
Check all plumbing fitting connections for leakage. Tighten or replace fittings. |

Error Code Display Screen | Cause | Action |
---|---|---|
0746-034C: Chamber temperature low during Sterilizing Mode
Instruments not sterile and need to be reprocessed. |
Maintenance not performed. |
Verify weekly and monthly maintenance has been performed by customer. Refer to User Guide. Perform monthly maintenance. |
Sterilizer is unlevel. |
Verify support surface is level. Remove tray/trays. Verify water on all sides of chamber is equal. |
|
Temperature sensor malfunctioning. Chamber temperature fell below lower sterilization temperature limit during Sterilization mode. |
Check for residue buildup on temperature sensor probe. Clean with Speed-Clean, distilled water, and abrasive pad. |
|
Verify that nothing is touching or crowding the temperature sensor (e.g., trays, pouches, cassettes). |
||
Perform Temperature Sensor Resistance Test. Allow sensor to cool to room temperature before testing. Acceptable range: 1.07 to 1.1K Ω Refer to: Temperature Sensor Resistance Test Replace temperature sensor. |
||
Perform Temperature Sensor Supply Voltage Test. Disconnect sensor harness from J6 on PC board. On pins 1 and 2, acceptable range is 4.0 to 6.0 VDC. Refer to: Temperature Sensor Supply Voltage Test Replace main PC board. |
||
Pressure leaks. |
Check for loose/worn door and dam gaskets or water leaking around door. Verify door and dam gaskets are installed properly. Replace gasket kit 1x year. Clean/replace gaskets. |
|
Check for voltage flow through harness. Refer to:Fill / Vent Valve Supply Voltage Test Replace harness. |
||
Leaking or defective vent valve: Check for water leaking from condensing coil spout. Perform Vent Valve Test. Clean/replace vent valve. |
||
Leaking or defective fill valve: Check for water leaking back into reservoir through the fill line. Look for bubbles coming from bottom of reservoir. Perform Fill Valve Test. Clean/replace fill valve. |
||
Leaking or defective air valve: Check for excessive steam coming through condensing coil spout. During the HEATING and VENTING modes, it is normal for steam to be exhausted through the spout. Clean/replace air valve. |
||
Leaking or defective Pressure Relief Valve: Check for water/steam leakage from beneath the rear of the sterilizer. Perform Pressure Relief Valve Test. Replace pressure relief valve. |
||
Pressure transducer tubing leaking: Check for steam leakage at pressure transducer tubing connections. Secure pressure transducer tubing connections with high temperature cable ties. |
||
Check all plumbing fitting connections for leakage. Tighten or replace fittings. |
||
Heating element malfunctioning. |
Inspect heating element. Heating element should lay flat in the bottom of the chamber and be free of any flaking/pitting. Perform Heating Element Resistance Test (should read 9 to 11 Ωfor 115V model and 34 to 42 Ω for 230V model). Replace heating element. |
|
PC board malfunctioning. |
Perform Heating Element Supply Voltage Test. With harness attached to leads, should have line voltage in test mode. |
|
Perform Main PC Board Relay Test. Check TP1 and TP2 for an acceptable range of 10 to 14 VDC. |
||
Perform Main PC Board Pressure Transducer Voltage Test. Acceptable range: 4.0 to 6.0 VDC |
||
Replace Main PC board. |

Error Code Display Screen | Cause | Action |
---|---|---|
0746-044C: Chamber temperature high during Sterilizing Mode
Instruments not sterile and need to be reprocessed. |
Maintenance not performed. |
Verify weekly and monthly maintenance has been performed by customer. Refer to User Guide. Perform monthly maintenance. |
Sterilizer is unlevel.
|
Verify support surface is level. Remove tray/trays. Verify water on all sides of chamber is equal. |
|
Temperature sensor malfunctioning. Chamber temperature rose above upper sterilization temperature limit during Sterilization mode. |
Check for residue buildup on temperature sensor probe. Clean with Speed-Clean, distilled water, and abrasive pad. |
|
Verify that nothing is touching or crowding the temperature sensor (e.g., trays, pouches, cassettes). |
||
Perform Temperature Sensor Resistance Test. Allow sensor to cool to room temperature before testing. Acceptable range: 1.07 to 1.1K Ω Refer to: Temperature Sensor Resistance Test Replace temperature sensor. |
||
Perform Temperature Sensor Supply Voltage Test. Disconnect sensor harness from J6 on PC board. On pins 1 and 2, acceptable range is 4.0 to 6.0 VDC. Refer to: Temperature Sensor Supply Voltage Test Replace main PC board. |
||
Not enough water. |
Check diameter of chamber water level sensor. Replace if .750 inch diameter. |
|
Pressure leaks. |
Check for loose/worn door and dam gaskets or water leaking around door. Verify door and dam gaskets are installed properly. Replace gasket kit 1x year. Clean/replace gaskets. |
|
Check for voltage flow through harness. Refer to:Fill / Vent Valve Supply Voltage Test Replace harness. |
||
Leaking or defective vent valve: Check for water leaking from condensing coil spout. Perform Vent Valve Test. Clean/replace vent valve. |
||
Leaking or defective fill valve: Check for water leaking back into reservoir through the fill line. Look for bubbles coming from bottom of reservoir. Perform Fill Valve Test. Clean/replace fill valve. |
||
Leaking or defective air valve: Check for excessive steam coming through condensing coil spout. During the HEATING and VENTING modes, it is normal for steam to be exhausted through the spout. Clean/replace air valve. |
||
Leaking or defective Pressure Relief Valve: Check for water/steam leakage from beneath the rear of the sterilizer. Perform Pressure Relief Valve Test. Replace pressure relief valve. |
||
Pressure transducer tubing leaking: Check for steam leakage at pressure transducer tubing connections. Secure pressure transducer tubing connections with high temperature cable ties. |
||
Check all plumbing fitting connections for leakage. Tighten or replace fittings. |
||
Heating element malfunctioning. |
Inspect heating element. Heating element should lay flat in the bottom of the chamber and be free of any flaking/pitting. Perform Heating Element Resistance Test (should read 9 to 11 Ω for 115V model and 34 to 42 Ω for 230V model). Replace heating element. |
|
PC board malfunctioning. |
Perform Heating Element Supply Voltage Test. With harness attached to leads, should have line voltage in test mode. |
|
Perform Main PC Board Relay Test. Check TP1 and TP2 for an acceptable range of 10 to 14 VDC. |
||
Perform Main PC Board Pressure Transducer Voltage Test. Acceptable range: 4.0 to 6.0 VDC |
||
Verify unit is operating within designated program parameters. Refer to: Using a Pressure Gauge | ||
Replace Main PC board. |

Error Code Display Screen | Cause | Action |
---|---|---|
0747-07[m]C: Chamber pressure sensor malfunction during [m] Mode
Modes 2, 3, 4: Instruments not sterile and need to be reprocessed. Modes 5, 6, 7: Risk of recontaminated instruments. Any other Mode: [not applicable] |
Chamber pressure reported as less than -5 PSI or greater than 43 PSI. |
Perform Air Valve Test. Clean/replace air valve. |
Power was lost for a few seconds during Filling, Heating, or Sterilizing mode (e.g., could be due to an electrical storm, brownout). |
Secure power cord connections at wall outlet and sterilizer receptacle. Unplug unit for 60 seconds; then plug into a dedicated, properly rated circuit. Check supply voltage. Contact an electrician to modify voltage. |
|
Operator may have unplugged unit when a previous error code was displayed. |
Check previous error codes. Refer to: Service History Errors Menu Refer to the error code(s) in this troubleshooting topic. |
|
Chamber filter clogged. |
Clean/replace filter. Clean with Speed-Clean and brush. |
|
Vent valve. |
Inspect vent valve. With the sterilizer in Venting mode, perform Vent Valve Testing. The valve is normally open and should not be energized during the Venting mode. Clean/replace vent valve. |
|
Temperature sensor malfunctioning. |
Check for residue buildup on temperature sensor probe. Clean with Speed-Clean, distilled water, and abrasive pad. |
|
Verify that nothing is touching or crowding the temperature sensor (e.g., trays, pouches, cassettes). |
||
Perform Temperature Sensor Resistance Test. Allow sensor to cool to room temperature before testing. Acceptable range: 1.07 to 1.1K Ω Refer to: Temperature Sensor Resistance Test Replace temperature sensor. |
||
Perform Temperature Sensor Supply Voltage Test. Disconnect sensor harness from J6 on PC board. On pins 1 and 2, acceptable range is 4.0 to 6.0 VDC. Refer to: Temperature Sensor Supply Voltage Test Replace main PC board. |
||
Pressure tubing at incorrect angle. |
Check tubing for restrictions and ensure tubing is angled down and away from the PC Board. Adjust pressure tubing to correct angle. |
|
PC board malfunctioning. |
Perform Heating Element Supply Voltage Test. With harness attached to leads, should have line voltage in test mode. |
|
Perform Main PC Board Relay Test. Check TP1 and TP2 for an acceptable range of 10 to 14 VDC. |
||
Perform Main PC Board Pressure Transducer Voltage Test. Acceptable range: 4.0 to 6.0 VDC |
||
Replace Main PC board. | ||
Facility voltage too high/low. |
Test facility supply voltage. Voltage must be: • 115V models: 104 to 127 VAC 50/60 Hz • 230V models: 207 to 253 VAC 50/60 Hz Contact an electrician to modify voltage. |
|
*[m] Modes: 0 = Power Up, 1 = Ready/Idle/Select Cycle, 2 = Filling, 3 = Heating, 4 = Sterilizing, 5 = Venting, 6 = Door Opening, 7 = Drying |

Error Code Display Screen | Cause | Action |
---|---|---|
0747-06[m]C: Chamber pressure high during [m] Mode
Modes 2, 3, 4: Instruments not sterile and need to be reprocessed. Modes 5, 6, 7: Risk of recontaminated instruments. Any other Mode: [not applicable] |
Chamber pressure reported as greater than 43 PSI. |
Perform Air Valve Test. Clean/replace air valve. |
Power was lost for a few seconds during Filling, Heating, or Sterilizing mode (e.g., could be due to an electrical storm, brownout). |
Secure power cord connections at wall outlet and sterilizer receptacle. Unplug unit for 60 seconds; then plug into a dedicated, properly rated circuit. Check supply voltage. Contact an electrician to modify voltage. |
|
Operator may have unplugged unit when a previous error code was displayed. |
Check previous error codes. Refer to: Access Troubleshooting Menu Refer to the error code(s) in this troubleshooting topic. |
|
Chamber filter clogged. |
Clean/replace filter. Clean with Speed-Clean and brush. |
|
Vent valve. |
Inspect vent valve. Clean valve, stretch spring. With the sterilizer in Venting mode, perform Vent Valve Testing. The valve is normally open and should not be energized during theVenting mode. Clean/replace vent valve. |
|
Pressure tubing at incorrect angle. |
Check tubing for restrictions and ensure tubing is angled down and away from the PC Board. Adjust pressure tubing to correct angle. |
|
Leaking/defective air valve. |
Perform Air Valve Test. Check for excessive steam coming through condensing coil spout. Clean/replace air valve. |
|
Sterilizer overloaded. |
Check the size of the load. Refer to User Guide. Reduce load size. |
|
Sterilizer is unlevel. |
Verify support surface is level. Remove tray/trays. Verify water on all sides of chamber is equal. |
|
Temperature sensor malfunctioning. |
Check for residue buildup on temperature sensor probe. Clean with Speed-Clean, distilled water, and abrasive pad. |
|
Verify that nothing is touching or crowding the temperature sensor (e.g., trays, pouches, cassettes). |
||
Perform Temperature Sensor Resistance Test. Allow sensor to cool to room temperature before testing. Acceptable range: 1.07 to 1.1K Ω Refer to: Temperature Sensor Resistance Test Replace temperature sensor. |
||
Perform Temperature Sensor Supply Voltage Test. Disconnect sensor harness from J6 on PC board. On pins 1 and 2, acceptable range is 4.0 to 6.0 VDC. Refer to: Temperature Sensor Supply Voltage Test Replace main PC board. |
||
PC board malfunctioning. |
Perform Heating Element Supply Voltage Test. With harness attached to leads, should have line voltage in test mode. |
|
Perform Main PC Board Relay Test. Check TP1 and TP2 for an acceptable range of 10 to 14 VDC. |
||
Perform Main PC Board Pressure Transducer Voltage Test. Acceptable range: 4.0 to 6.0 VDC |
||
Replace Main PC board. | ||
Blockage or malfunction in VistaCool System (if used). |
Check for debris or blockage in the tubing/fittings to the VistaCool System. Clean/replace as required. |
|
Check operation of In-Iine Thermal Sensor. Replace Thermal Regulator Wax Motor or Thermal Sensor. |
||
*[m] Modes: 0 = Power Up, 1 = Ready/Idle/Select Cycle, 2 = Filling, 3 = Heating, 4 = Sterilizing, 5 = Venting, 6 = Door Opening, 7 = Drying |

Error Code Display Screen | Cause | Action |
---|---|---|
0747-0428: Chamber pressure high during Filling Mode
Instruments not sterile and need to be reprocessed. |
Chamber pressure > 15 kPa during Fill Mode. |
Perform Air Valve Test. Clean/replace air valve. |
Leaking or defective air valve. |
Perform Air Valve Test. Check for excessive steam coming through condensing coil spout. Clean/replace air valve. |
|
Restricted condensing coil. |
Check for blockage in coil and ensure weep hole is present and not obstructed. Clear blockage / replace condensing coil. |
|
Pressure tubing at incorrect angle. |
Check tubing for restrictions and ensure tubing is angled down and away from the PC Board. Adjust pressure tubing to correct angle. |
|
Blockage or malfunction in VistaCool System (if used). |
Check for debris or blockage in the tubing/fittings to the VistaCool System. Clean/replace as required. |
|
Check operation of In-Iine Thermal Sensor. Replace Thermal Regulator Wax Motor or Thermal Sensor. |
||
Tank is lower than the facility drain. |
Install the tank higher than the facility drain. |
|
PC board malfunctioning. |
Perform Heating Element Supply Voltage Test. With harness attached to leads, should have line voltage in test mode. |
|
Perform Main PC Board Relay Test. Check TP1 and TP2 for an acceptable range of 10 to 14 VDC. |
||
Perform Main PC Board Pressure Transducer Voltage Test. Acceptable range: 4.0 to 6.0 VDC |
||
Replace Main PC board. |

Error Code Display Screen | Cause | Action |
---|---|---|
0747-033C: Chamber pressure low during Heating Mode
Instruments not sterile and need to be reprocessed. |
Maintenance not performed. |
Verify weekly and monthly maintenance has been performed by customer. Refer to User Guide. Perform monthly maintenance. |
Sterilizer is unlevel. |
Verify support surface is level. Remove tray/trays. Verify water on all sides of chamber is equal. |
|
If the unit is building temperature... |
||
Temperature sensor malfunctioning. The first air purge did not occur within 40 minutes of the start of Heat Up mode. |
Check for residue buildup on temperature sensor probe. Clean with Speed-Clean, distilled water, and abrasive pad. |
|
Verify that nothing is touching or crowding the temperature sensor (e.g., trays, pouches, cassettes). |
||
Perform Temperature Sensor Resistance Test. Allow sensor to cool to room temperature before testing. Acceptable range: 1.07 to 1.1K Ω Refer to: Temperature Sensor Resistance Test Replace temperature sensor. |
||
Perform Temperature Sensor Supply Voltage Test. Disconnect sensor harness from J6 on PC board. On pins 1 and 2, acceptable range is 4.0 to 6.0 VDC. Refer to: Temperature Sensor Supply Voltage Test Replace main PC board. |
||
Pressure leaks. |
Check for loose/worn door and dam gaskets or water leaking around door. Verify door and dam gaskets are installed properly. Replace gasket kit 1x year. Clean/replace gaskets. |
|
Check for voltage flow through harness. Refer to:Fill / Vent Valve Supply Voltage Test Replace harness. |
||
Leaking or defective vent valve: Check for water leaking from condensing coil spout. Perform Vent Valve Test. Clean/replace vent valve. |
||
Leaking or defective fill valve: Check for water leaking back into reservoir through the fill line. Look for bubbles coming from bottom of reservoir. Perform Fill Valve Test. Clean/replace fill valve. |
||
Leaking or defective air valve: Check for excessive steam coming through condensing coil spout. During the HEATING and VENTING modes, it is normal for steam to be exhausted through the spout. Clean/replace air valve. |
||
Leaking or defective Pressure Relief Valve: Check for water/steam leakage from beneath the rear of the sterilizer. Perform Pressure Relief Valve Test. Replace pressure relief valve. |
||
Pressure transducer tubing leaking: Check for steam leakage at pressure transducer tubing connections. Secure pressure transducer tubing connections with high temperature cable ties. |
||
Check all plumbing fitting connections for leakage. Tighten or replace fittings. |
||
If the unit is not building temperature... |
||
Heating element malfunctioning. |
Inspect heating element. Heating element should lay flat in the bottom of the chamber and be free of any flaking/pitting. Perform Heating Element Resistance Test (should read 9 to 11 Ω for 115V model and 34 to 42 Ω for 230V model). Replace heating element. |
|
Inspect heating element wires for corrosion or if disconnected. Reconnect/replace heating element wires. |
||
PC board malfunctioning. |
Perform Heating Element Supply Voltage Test. With harness attached to leads, should have line voltage in test mode. |
|
Perform Main PC Board Relay Test. Check TP1 and TP2 for an acceptable range of 10 to 14 VDC. |
||
Perform Main PC Board Pressure Transducer Voltage Test. Acceptable range: 4.0 to 6.0 VDC |
||
Replace Main PC board. |

Error Code Display Screen | Cause | Action |
---|---|---|
0747-043C: Chamber pressure high during Heating Mode
Instruments not sterile and need to be reprocessed. |
The chamber pressure did not drop during an air purge. |
Perform Air Valve Test. Clean/replace air valve. |
Sterilizer overloaded. |
Check the size of the load. Refer to User Guide. Reduce load size. |
|
Chamber filter clogged. |
Clean/replace filter. Clean with Speed-Clean and brush. |
|
Maintenance not performed. |
Verify weekly and monthly maintenance has been performed by customer. Refer to User Guide. Perform monthly maintenance. |
|
Leaking or defective air valve. |
Perform Air Valve Test. Check for excessive steam coming through condensing coil spout. Clean/replace air valve. |
|
Pressure tubing at incorrect angle. |
Check tubing for restrictions and ensure tubing is angled down and away from the PC Board. Adjust pressure tubing to correct angle. |
|
PC board malfunctioning. |
Perform Heating Element Supply Voltage Test. With harness attached to leads, should have line voltage in test mode. |
|
Perform Main PC Board Relay Test. Check TP1 and TP2 for an acceptable range of 10 to 14 VDC. |
||
Perform Main PC Board Pressure Transducer Voltage Test. Acceptable range: 4.0 to 6.0 VDC |
||
Replace Main PC board. | ||
Blockage or malfunction in VistaCool System (if used). |
Check for debris or blockage in the tubing/fittings to the VistaCool System. Clean/replace as required. |
|
Check operation of In-Iine Thermal Sensor. Replace Thermal Regulator Wax Motor or Thermal Sensor. |

Error Code Display Screen | Cause | Action |
---|---|---|
0747-0458: Chamber pressure high during Venting Mode
Risk of recontaminated instruments. |
The chamber pressure is greater than 5 kPa for 10 minutes in fast vent and for 60 minutes in slow vent |
Inspect vent valve. Clean valve, stretch spring. Perform Vent Valve Testing. Clean/replace vent valve. |
Maintenance not performed. |
Verify weekly and monthly maintenance has been performed by customer. Refer to User Guide. Perform monthly maintenance. |
|
Chamber filter clogged. |
Clean/replace filter. Clean with Speed-Clean and brush. |
|
Vent valve. |
Inspect vent valve. Clean valve, stretch spring. With the sterilizer in Venting mode, perform Vent Valve Testing. The valve is normally open and should not be energized during the Venting mode. Clean/replace vent valve. |
|
Pressure tubing at incorrect angle. |
Check tubing for restrictions and ensure tubing is angled down and away from the PC Board. Adjust pressure tubing to correct angle. |
|
PC board malfunctioning. |
Perform Heating Element Supply Voltage Test. With harness attached to leads, should have line voltage in test mode. |
|
Perform Main PC Board Relay Test. Check TP1 and TP2 for an acceptable range of 10 to 14 VDC. |
||
Perform Main PC Board Pressure Transducer Voltage Test. Acceptable range: 4.0 to 6.0 VDC |
||
Replace Main PC board. | ||
Blockage or malfunction in VistaCool System (if used). |
Check for debris or blockage in the tubing/fittings to the VistaCool System. Clean/replace as required. |
|
Check operation of In-Iine Thermal Sensor. Replace Thermal Regulator Wax Motor or Thermal Sensor. |
||
Tank is lower than the facility drain. |
Install the tank higher than the facility drain. |

Error Code Display Screen | Cause | Action |
---|---|---|
0747-0468: Chamber pressure high during Door Opening Mode
Risk of recontaminated instruments. |
The chamber pressure is greater than 5 kPa during Door Open mode. |
Check for residue buildup on temperature sensor probe. Clean with Speed-Clean, distilled water, and abrasive pad. |
Perform Temperature Sensor Resistance Test. Allow sensor to cool to room temperature before testing. Acceptable range: 1.07 to 1.1K Ω Refer to: Temperature Sensor Resistance Test Replace temperature sensor. |
||
Perform Temperature Sensor Supply Voltage Test. Disconnect sensor harness from J6 on PC board. On pins 1 and 2, acceptable range is 4.0 to 6.0 VDC. Refer to: Temperature Sensor Supply Voltage Test Replace main PC board. |
||
Chamber filter clogged. |
Clean/replace filter. Clean with Speed-Clean and brush. |
|
Leaking or defective air valve. |
Perform Air Valve Test. Check for excessive steam coming through condensing coil spout. Clean/replace air valve. |
|
Vent valve. |
Inspect vent valve. Clean valve, stretch spring. With the sterilizer in Venting mode, perform Vent Valve Testing. The valve is normally open and should not be energized during the Venting mode. Clean/replace vent valve. |

Display / Symptom | Cause | Action |
---|---|---|
0747-0478: Chamber pressure high during Drying Mode
Risk of recontaminated instruments. |
The chamber pressure is greater than 5 kPa during Drying mode. |
If door does not open, check door switches and door motor. Check tubing routing and connections to pressure sensor. Remove power for 10 seconds. Plug back in and see if error clears. If not, replace main control PCB. |

Display / Symptom | Cause | Action |
---|---|---|
0749-1568: Sensor not detecting door opening
Risk of recontaminated instruments. |
The door motor hall effect sensor did not detect the door latch lever magnet after energizing the door motor. |
Check door motor operation. Check connections to the door motor hall effect sensor. Check door motor cable harness. Remove power for 10 seconds. Plug back in and see if error clears. If not, replace main control PCB. |

Error Code Display Screen | Cause | Action |
---|---|---|
074A-08[m]C*: Thermostat open during [m] Mode
Modes 2, 3, 4: Instruments not sterile and need to be reprocessed. Modes 5, 6, 7: Risk of recontaminated instruments. Any other Mode: [not applicable] |
Maintenance not performed. |
Verify weekly and monthly maintenance has been performed by customer. Refer to User Guide. Perform monthly maintenance. |
Sterilizer was unplugged and then plugged back in while thermostat was still tripped. |
Verify if error was preceded with other error codes (e.g., 074A-08[m]C, 0746-06[m]C). Unplug unit for 60 seconds; then plug into a dedicated, properly rated circuit. Allow unit to cool for 30 minutes. |
|
High Limit Thermostats open/ malfunctioning. |
Verify if unit still overheated. Unplug unit for 60 seconds; then plug it into a dedicated, properly rated circuit. Allow unit to cool for 30 minutes. If error code does not reset, perform Thermostat Continuity Test. Meter Reading — Anything but OL. Replace thermostats. |
|
Temperature sensor malfunctioning. |
Check for residue buildup on temperature sensor probe. Clean with Speed-Clean, distilled water, and abrasive pad. |
|
Verify that nothing is touching or crowding the temperature sensor (e.g., trays, pouches, cassettes). |
||
Perform Temperature Sensor Resistance Test. Allow sensor to cool to room temperature before testing. Acceptable range: 1.07 to 1.1K Ω Refer to: Temperature Sensor Resistance Test Replace temperature sensor. |
||
Perform Temperature Sensor Supply Voltage Test. Disconnect sensor harness from J6 on PC board. On pins 1 and 2, acceptable range is 4.0 to 6.0 VDC. Refer to: Temperature Sensor Supply Voltage Test Replace main PC board. |
||
Loose/damaged wire connection. |
Check wire connections at high-limit thermostat and at J15 on Main PC Board. Secure/repair loose or damaged wire connections. |
|
Door hinge binding. |
Check for binding or buildup of debris in the door hinges. Clean debris from hinges. |
|
Facility voltage too high/low. |
Test facility supply voltage. Voltage must be: • 115V models: 104 to 127 VAC 50/60 Hz • 230V models: 207 to 253 VAC 50/60 Hz Contact an electrician to modify voltage. |
|
If the unit skips the Filling mode... |
||
Tray rack / tray plate not installed properly. |
Install tray rack / tray plate properly. The 45 degree angled end of tray plate must be positioned upward and to the back of the chamber. |
|
(M11 only) Mesh filter not in place. |
If filter is not in place, re-install filter. | |
Office running back-to-back cycles. |
Check if operator is running consecutive cycles without completing the DRY cycle. Complete full cycle. |
|
Water level sensor malfunctioning. Dirty/foreign object touching sensor. |
Inspect water level sensor for foreign objects. burs, instruments, etc. Remove foreign objects. Clean/dry water level sensor. Clean with Speed-Clean and Scotch-Brite pad. |
|
Perform Water Level Sensor Test. Replace water level sensor. |
||
If the unit completes the Filling mode… |
||
Sterilizer is unlevel. |
Verify support surface is level. Remove tray/trays. Verify water on all sides of chamber is equal. |
|
Verify that nothing is touching or crowding the temperature sensor (e.g., trays, pouches, cassettes). |
||
Not enough water. |
Check diameter of chamber water level sensor. Replace if .750 inch diameter. |
|
Pressure leaks. |
Check for loose/worn door and dam gaskets or water leaking around door. Verify door and dam gaskets are installed properly. Replace gasket kit 1x year. Clean/replace gaskets. |
|
Check for voltage flow through harness. Refer to:Fill / Vent Valve Supply Voltage Test Replace harness. |
||
Leaking or defective vent valve: Check for water leaking from condensing coil spout. Perform Vent Valve Test. Clean/replace vent valve. |
||
Leaking or defective fill valve: Check for water leaking back into reservoir through the fill line. Look for bubbles coming from bottom of reservoir. Perform Fill Valve Test. Clean/replace fill valve. |
||
Leaking or defective air valve: Check for excessive steam coming through condensing coil spout. During the HEATING and VENTING modes, it is normal for steam to be exhausted through the spout. Clean/replace air valve. |
||
Leaking or defective Pressure Relief Valve: Check for water/steam leakage from beneath the rear of the sterilizer. Perform Pressure Relief Valve Test. Replace pressure relief valve. |
||
Pressure transducer tubing leaking: Check for steam leakage at pressure transducer tubing connections. Secure pressure transducer tubing connections with high temperature cable ties. |
||
Check all plumbing fitting connections for leakage. Tighten or replace fittings. |
||
Heating element malfunctioning. |
Inspect heating element. Heating element should lay flat in the bottom of the chamber and be free of any flaking/pitting. Perform Heating Element Resistance Test (should read 9 to 11 Ω for 115V model and 34 to 42 Ω for 230V model). Replace heating element. |
|
PC board malfunctioning. |
Perform Heating Element Supply Voltage Test. With harness attached to leads, should have line voltage in test mode. |
|
Perform Main PC Board Relay Test. Check TP1 and TP2 for an acceptable range of 10 to 14 VDC. |
||
Perform Main PC Board Pressure Transducer Voltage Test. Acceptable range: 4.0 to 6.0 VDC |
||
Replace Main PC board. | ||
Sterilizer overloaded. |
Check the size of the load. Refer to User Guide. Reduce load size. |
|
*[m] Modes: 0 = Power Up, 1 = Ready/Idle/Select Cycle, 2 = Filling, 3 = Heating, 4 = Sterilizing, 5 = Venting, 6 = Door Opening, 7 = Drying |

Display / Symptom | Cause | Action |
---|---|---|
074F-4314: Incorrect parameter applied to cycle |
Communication error |
Check communication cable connections between circuit boards. Remove power for 10 seconds; plug back in and see if error clears. If not, replace main control PCB. |

Display / Symptom | Cause | Action |
---|---|---|
0751-4468: Door not detected in Dry position
Risk of recontaminated instruments. |
Door position switch not engaged in Door Opening mode. |
Check door dry position switch operation and connections to control board. Remove power for 10 seconds; plug back in and see if error clears. If not, replace main control PCB. |

Display / Symptom | Cause | Action |
---|---|---|
0751-4478: Door not detected in Dry position
Risk of recontaminated instruments. |
Door position switch not engaged in Drying mode. |
Check door dry position switch operation and connections to control board. Remove power for 10 seconds; plug back in and see if error clears. If not, replace main control PCB. |

Display / Symptom | Cause | Action |
---|---|---|
0752-04[m]8: Board temperature sensor too high in [m] mode
Modes 2, 3, 4: Instruments not sterile and need to be reprocessed. Modes 5, 6, 7: Risk of recontaminated instruments. Any other Mode: [not applicable] |
Altimeter temperature reported as above 80 °C. |
Remove power for at least five minutes. Plug back in and see if the error clears. If not, replace PCB. |
*[m] Modes: 0 = Power Up, 1 = Ready/Idle/Select Cycle, 2 = Filling, 3 = Heating, 4 = Sterilizing, 5 = Venting, 6 = Door Opening, 7 = Drying |

Display / Symptom | Cause | Action |
---|---|---|
0709-0F08: CONTROL BOARD ERROR Software error during Power Up |
No valid application image found in either bank on power up. |
Update firmware, if available. Remove power for 10 seconds. Plug back in and see if error clears. If not, replace main control PCB. |