PowerVac G Troubleshooting Chart
When working on Twin units, unplug parallel harness.

Display/Symptom |
Cause |
Check |
Action |
---|---|---|---|
N/A |
No power. |
Check on/off switch on front of the base. |
Verify on/off switch is in the on postion. |
Check disconnect box/ main circuit breaker. |
Verify disconnect box/main circuit breaker is in the on postion. |
||
Low Voltage Harness disconnected/loose. |
Check door limit switch, on/off switch, and master control panel harness. |
Secure/Replace harness connections. |
|
Faulty Switch |
Check LED light “D4”, if “ON” only check Low Voltage. If LED light is not on, preform checks on Door Limit switch, On/Off switch, and master control panel. Refer to: PC Board to check LED light, Door Limit Switch, On/Off Switch,and Master Control Panel. |
Replace Switch |
|
Separator tank not draining. |
Possible Clog in tank. Remove lid to see if tank is full. |
Remove clog. |
|
Check valve is stuck closed. Refer to: Check valve. |
Repair/ Replace check valve |
||
Turn power off and unplug float harness from base unit. Turn unit back on. If the unit runs, check Float |
Replace Float Assembly |
||
Wire harness between VFD and PC board loose or broken. |
Check for loose or broken wire harness between VFD and PC board. Ensure grey cable is plugged into VFD and PC board. |
Secure/Replace harness connections. |
|
Pump is locked up. |
Feel for pump pulley drag or sticking. Refer to: Pump. |
Contact Midmark Technical Service. 1-800-643-6275 |
|
PC board Malfunctioning. |
Turn main power supply on. Keep unit turned off. Unplug wire harness from J6 on PC board. Test resistance. Set meter to Ω. Meter Reading > 1 Ω, Refer to: PC Board |
Replace Main PC Board |
|
Check Daughter Card. Refer to: Daughter Card |
Replace Main PC Board |
||
Contactor not working properly. |
Check Contactor. Refer to: Contactor |
Replace Contactor. |
|
VFD not working properly. |
Check if there is an error code displayed. Refer to: VFD Error Code Chart |
Follow required actions in testing. |
|
Lights not working on PC board, check Fuse. Refer to: Fuse and Transformer. |
Replace Fuse |
||
Lights not working on VFD, check Line Filter. Refer to: Line Filter |
Secure/Replace harness connections. |
||
Lights not working on VFD, check Contactor. Refer to: Contactor |
Replace Contactor |
||
Lights not working on PC board, check transformer. Refer to: Fuse and Transformer |
Replace Transformer |
||
Lights not working on VFD. Check Ohms at J6 terminal. If greater than 5 ohm, Refer to: PC Board |
Replace PC Board |
||
Lights not working on VFD. Disconnect drive harness at J6 Location to check for 24 VDC. If Voltage is not present, Ohm out drive harness. Refer to: VFD Test. |
Replace Drive Harness |
||
Lights not working on VFD. Check for line Voltage (208-230V) at incoming terminal strip on VFD. If Line voltage is present, Refer to: Variable Frequency Drive |
Replace VFD |
||
Motor bad. |
Perform all motor checks. Refer to: Motor and Pump |
Replace Motor |

Display/Symptom |
Cause |
Check |
Action |
---|---|---|---|
N/A |
Motor not running. |
||
Leaking or loose connection. |
Check hoses and bulkhead fittings. |
Secure/Replace hoses or bulkhead connections. |
|
Check Separator lid gasket. |
Replace Separator Lid Gasket |
||
Check drain line check valve. Refer to: Check Valve |
Replace Check Valve. |
||
Pump pulley not spinning freely. |
Moisture in the pump. |
P trap not installed properly. |
Drain, release pump, install missing P trap per instructions. |
P trap drain system clogged. |
Drain, release pump, remove clog from P trap drain. |
||
Incorrect line cleaners. |
Customer should be using non sudsing line cleaners. |
Release pump, Refer to: User’s Manual. |
|
Incorrect line cleaning process. |
Check how customer is cleaning lines. |
Release pump, Refer to: User’s Manual. |
|
Suds from ultrasonic cleaner getting into pump. |
Suction line close to ultrasonic. Inquire with staff. |
Release pump, advise customer not to use product to drain ultrasonic cleaner. |
|
Timer Kit not installed properly. |
Installation of kit. |
Release pump, correct or install timer kit. |
|
Timer kit faulty |
Check dip switch location on timer board. Dip switches are to be in down position or timer is malfunctioning. |
Release pump, repair, replace timer kit. |
|
No power to timer kit. |
If power is being turned off to vacuum system. Vacuum system must be powered for timer to function correctly. |
Release pump, continual power needed for timer kit. |
|
N/A |
Exhaust Hose Clogged |
Remove exhaust hose to see if suction increases. If suction increased, check for clog. |
Unclog or replace exhaust line |
Pressure sensor not working properly. |
Check pressure sensor. Refer to: Pressure Sensor |
Replace pressure sensor. |

Display/Symptom |
Cause |
Check |
Action |
---|---|---|---|
N/A |
Ambient temperature too high. |
Check Equipment Room Temperature, should be 40° to 104° Fahrenheit or 4° to 37° Celsius. |
Refer to: Installation Manual site requirements |
Contactor not working properly. |
Check Contactor. Refer to: Contactor |
Replace Contactor |
|
Pump is locked up. |
Refer to: Pump |
Contact Midmark Technical Service. 1-800-643-6275 |
|
Exhaust Hose Clogged |
Remove exhaust hose to see if suction increases. If suction increases, check for clog. |
Unclog or replace exhaust line |
|
Motor not running. |

Display/Symptom |
Cause |
Check |
Action |
---|---|---|---|
N/A |
Pressure sensor not working properly. |
Check pressure sensor. Refer to: Pressure Sensor |
Replace Pressure sensor |
Plumbing lines leaking. |
Check office plumbing system for leaks. |
Repair/Replace Plumbing Lines |
|
VRV Malfunctioning |
Check VRV. Refer to: VRV |
Adjust/Replace VRV |

Display/Symptom |
Cause |
Check |
Action |
---|---|---|---|
Leaking Oil |
Oil Reservoir is overfilled. |
Check sight glass. Refer to: Pump |
Drain Oil |
Pump is overheating |
Check ambient temperature in the room. Temperature should be 40° to 104° F or 4° to 40° C |
Contact a certified specialist to adjust temperature. |
|
Check Clearance around unit. Allow 12 inches on the front and right side and allow 6 inches on the left side and back |
Adjust base unit to clearance specifications. |
||
Check if fans are working. Refer to: Fans |
Replace fans |
||
Minor Oil leak |
Minimal oil leakage may occur with no harm to the vacuum. Refer to the oil sight glass fill level requirements. |
Monitor oil levels and fill as needed. Replace absorbent pad if required. |